Title: Technical Note: Experimental results from a prototype high-field inline MRI-linac

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4961395· OSTI ID:22689287
;  [1];
  1. Department of Medical Physics, Ingham Institute for Applied Medical Research, Liverpool NSW 2170 (Australia)

Purpose: The pursuit of real-time image guided radiotherapy using optimal tissue contrast has seen the development of several hybrid magnetic resonance imaging (MRI)-treatment systems, high field and low field, and inline and perpendicular configurations. As part of a new MRI-linac program, an MRI scanner was integrated with a linear accelerator to enable investigations of a coupled inline MRI-linac system. This work describes results from a prototype experimental system to demonstrate the feasibility of a high field inline MR-linac. Methods: The magnet is a 1.5 T MRI system (Sonata, Siemens Healthcare) was located in a purpose built radiofrequency (RF) cage enabling shielding from and close proximity to a linear accelerator with inline (and future perpendicular) orientation. A portable linear accelerator (Linatron, Varian) was installed together with a multileaf collimator (Millennium, Varian) to provide dynamic field collimation and the whole assembly built onto a stainless-steel rail system. A series of MRI-linac experiments was performed to investigate (1) image quality with beam on measured using a macropodine (kangaroo) ex vivo phantom; (2) the noise as a function of beam state measured using a 6-channel surface coil array; and (3) electron contamination effects measured using Gafchromic film and an electronic portal imaging device (EPID). Results: (1) Image quality was unaffected by the radiation beam with the macropodine phantom image with the beam on being almost identical to the image with the beam off. (2) Noise measured with a surface RF coil produced a 25% elevation of background intensity when the radiation beam was on. (3) Film and EPID measurements demonstrated electron focusing occurring along the centerline of the magnet axis. Conclusions: A proof-of-concept high-field MRI-linac has been built and experimentally characterized. This system has allowed us to establish the efficacy of a high field inline MRI-linac and study a number of the technical challenges and solutions.

OSTI ID:
22689287
Journal Information:
Medical Physics, Vol. 43, Issue 9; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

TH-AB-BRA-12: Experimental Results From the First High-Field Inline MRI-Linac
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22689287

Daily QA of linear accelerators using only EPID and OBI
Journal Article · Thu Oct 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22689287

Impact of the MLC on the MRI field distortion of a prototype MRI-linac
Journal Article · Sun Dec 15 00:00:00 EST 2013 · Medical Physics · OSTI ID:22689287