Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

A clinically observed discrepancy between image-based and log-based MLC positions

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4949002· OSTI ID:22685142
; ; ; ; ;  [1]
  1. Department of Radiation Oncology, University of Virginia, Charlottesville, Virginia 22908 (United States)
Purpose: To present a clinical case in which real-time intratreatment imaging identified an multileaf collimator (MLC) leaf to be consistently deviating from its programmed and logged position by >1 mm. Methods: An EPID-based exit-fluence dosimetry system designed to prevent gross delivery errors was used to capture cine during treatment images. The author serendipitously visually identified a suspected MLC leaf displacement that was not otherwise detected. The leaf position as recorded on the EPID images was measured and log-files were analyzed for the treatment in question, the prior day’s treatment, and for daily MLC test patterns acquired on those treatment days. Additional standard test patterns were used to quantify the leaf position. Results: Whereas the log-file reported no difference between planned and recorded positions, image-based measurements showed the leaf to be 1.3 ± 0.1 mm medial from the planned position. This offset was confirmed with the test pattern irradiations. Conclusions: It has been clinically observed that log-file derived leaf positions can differ from their actual position by >1 mm, and therefore cannot be considered to be the actual leaf positions. This cautions the use of log-based methods for MLC or patient quality assurance without independent confirmation of log integrity. Frequent verification of MLC positions through independent means is a necessary precondition to trust log-file records. Intratreatment EPID imaging provides a method to capture departures from MLC planned positions.
OSTI ID:
22685142
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-F-T-469: A Clinically Observed Discrepancy Between Image-Based and Log- Based MLC Position
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649059

EPID-based verification of the MLC performance for dynamic IMRT and VMAT
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22099044

Using an EPID for patient-specific VMAT quality assurance
Journal Article · Tue Mar 15 00:00:00 EDT 2011 · Medical Physics · OSTI ID:22096925