Title: TH-CD-207B-11: Multi-Vendor Phantom Study of CT Lung Density Metrics: Is a Reproducibility of Less Than 1 HU Achievable?

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4958217· OSTI ID:22679347
 [1];  [2];  [3];  [4];  [5];  [6]
  1. National Institute of Standards & Technology, Gaithersburg, MD (United States)
  2. Brigham & Women’s Hospital, Boston, MA (United States)
  3. University of Wisconsin, Madison, WI (United States)
  4. Toshiba Medical Research Institute USA, Inc, Vernon Hills, IL (United States)
  5. Nation Jewish Health, Denver, CO (United States)
  6. Siemens Medical Solutions USA, Inc., Baltimore, MD (United States)

Purpose: To standardize the calibration procedures of CT lung density measurements using low-density reference foams in a phantom, and to demonstrate a reproducibility of less than 1 HU for lung equivalent foam densities measured across CT vendor platforms and protocols. Methods: A phantom study was conducted on CT scanner models from 4 vendors at 100, 120, and 135/140 kVp and 1.5, 3, and 6 mGy dose settings, using a lung density phantom containing air, water, and 3 reference foams (indirectly calibrated) with discrete densities simulating a 5-cm slice of the human chest. Customized segmentation software was used to analyze the images and generate a mean HU and variance for each of the density for the 22 vendor/protocols. A 3-step calibration process was devised to remove a scanner-dependent parameter using linear regression of the HU value vs the relative electron density. The results were mapped to a single energy (80 keV) for final comparison. Results: The heterogeneity across vendor platforms for each density assessed by a random effects model was reduced by 50% after re-calibration, while the standard deviation of the mean HU values also improved by about the same amount. The 95% CI of the final HU value was within +/−1 HU for all 3 reference foam densities. For the backing lung foam in the phantom (served as an “unknown”), this CI is +/− 1.6 HU. The kVp and dose settings did not appear to have significant contributions to the variability. Conclusion: With the proposed calibration procedures, the inter-scanner reproducibility of better than 1 HU is demonstrated in the current phantom study for the reference foam densities, but not yet achieved for a test density. The sources of error are being investigated in the next round of scanning with a certified Standard Reference Material for direct calibration. Fain: research funding from GE Healthcare to develop pulmonary MRI techniques. Hoppel: employee of Toshiba Medical Research Institute USA/financial interest with GE Healthcare. M. Fuld: employee of Siemens Healthcare for medical device equipment and software. This project is supported partially by RSNA QIBA Concept Award (Fain), NIH/NIBIB, HHSN268201300071C (Y).

OSTI ID:
22679347
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

TH-A-207B-00: Shear-Wave Imaging and a QIBA US Biomarker Update
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22679347

TH-A-207B-01: Basics and Current Implementations of Ultrasound Imaging of Shear Wave Speed and Elasticity
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22679347

TH-A-207B-02: QIBA Ultrasound Elasticity Imaging System Biomarker Qualification and User Testing of Systems
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22679347