TH-CD-BRA-10: Towards Reference Dosimetry of MR-Linacs Using a Clinical Probe-Format Calorimeter
- McGill University, Montreal, QC (Canada)
- University Medical Center Utrecht, Utrecht (Netherlands)
Purpose: To evaluate the influence of a 1.5 T magnetic field (B-field) on the response of a small-scale graphite calorimeter probe (GPC) developed for use as a novel clinical reference dosimeter. Characterization of the GPC was also assessed in a hybrid MRI-linac (MRL) clinical prototype by performing absolute dosimetry in multiple detector orientations. Methods: B-field influence was characterized using a variable-strength electromagnet system located 280 cm from the source of a clinical linac. The GPC was used to perform a total of 160 absolute dose measurements (6 MV, 920 MU/min) in a water phantom placed between the poles of the electromagnet. The magnitude of the B-field between the poles was varied in the range of 0 – 1.5 T. The relative response of the GPC was determined and compared to that of a thimble type ionization chamber (Exradin A1SL, Standard Imaging). Next, 65 dose measurements were performed using the GPC in a clinical MRL field (7 MV, 620 MU/min) to quantify the rotational dependence of the detector in the presence of a 1.5 T B-field. The GPC was rotated in steps of 90° inside a graphite phantom (SSD 140 cm, depth 2.5 cm) for two detector orientations (parallel and perpendicular to the B field). Results: Relative to the zero B-field condition, the A1SL chamber exhibited an average overresponse of +1.2 % ± 0.03 % at a B-field of 1.5 T, while the GPC under-responded on average by −0.5 % ± 0.9 %. For the MRL measurements, no significant differences were observed between the parallel and perpendicular orientations. In both cases, a rotational dependence of approximately ±1 % was measured. Conclusion: This work suggests that the B-field has minimal influence on the response of the GPC, making it a potentially attractive solution for clinical MRL reference dosimetry. This work has been supported in part by the CREATE Medical Physics Research Training Network NSERC grant RGPIN 432290, as well as NSERC grants RGPIN 298191 & 435608. JR is a scholar from The Terry Fox Foundation Strategic Training Initiative for Excellence in Radiation Research for the 21st Century (EIRR21).
- OSTI ID:
- 22679295
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
WE-AB-BRB-01: Development of a Probe-Format Graphite Calorimeter for Practical Clinical Dosimetry: Numerical Design Optimization, Prototyping, and Experimental Proof-Of-Concept
SU-E-T-485: Investigation of a Synthetic Diamond Detector for Tomotherapy Dosimetry
Consequences of air around an ionization chamber: Are existing solid phantoms suitable for reference dosimetry on an MR-linac?
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22570078
SU-E-T-485: Investigation of a Synthetic Diamond Detector for Tomotherapy Dosimetry
Journal Article
·
Sun Jun 01 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22369627
Consequences of air around an ionization chamber: Are existing solid phantoms suitable for reference dosimetry on an MR-linac?
Journal Article
·
Fri Jul 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22689454