WE-FG-BRA-06: Systematic Study of Target Localization for Bioluminescence Tomography Guided Radiation Therapy for Preclinical Research

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4957906· OSTI ID:22679075
; ; ;  [1];  [1];  [2];  [3];  [3];  [4]
  1. Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, Baltimore, MD (United States)
  2. Laboratory for Computational Sensing and Robotics, Johns Hopkins University, Baltimore, MD (United States)
  3. Department of Oncology, Department of Surgery, Johns Hopkins University, Baltimore, MD (United States)
  4. McMaster University, Hamilton, Ontario, CA (Canada)
Purpose: To overcome the limitation of CT/CBCT in guiding radiation for soft tissue targets, we developed a bioluminescence tomography(BLT) system for preclinical radiation research. We systematically assessed the system performance in target localization and the ability of resolving two sources in simulations, phantom and in vivo environments. Methods: Multispectral images acquired in single projection were used for the BLT reconstruction. Simulation studies were conducted for single spherical source radius from 0.5 to 3 mm at depth of 3 to 12 mm. The same configuration was also applied for the double sources simulation with source separations varying from 3 to 9 mm. Experiments were performed in a standalone BLT/CBCT system. Two sources with 3 and 4.7 mm separations placed inside a tissue-mimicking phantom were chosen as the test cases. Live mice implanted with single source at 6 and 9 mm depth, 2 sources with 3 and 5 mm separation at depth of 5 mm or 3 sources in the abdomen were also used to illustrate the in vivo localization capability of the BLT system. Results: Simulation and phantom results illustrate that our BLT can provide 3D source localization with approximately 1 mm accuracy. The in vivo results are encouraging that 1 and 1.7 mm accuracy can be attained for the single source case at 6 and 9 mm depth, respectively. For the 2 sources study, both sources can be distinguished at 3 and 5 mm separations at approximately 1 mm accuracy using 3D BLT but not 2D bioluminescence image. Conclusion: Our BLT/CBCT system can be potentially applied to localize and resolve targets at a wide range of target sizes, depths and separations. The information provided in this study can be instructive to devise margins for BLT-guided irradiation and suggests that the BLT could guide radiation for multiple targets, such as metastasis. Drs. John W. Wong and Iulian I. Iordachita receive royalty payment from a licensing agreement between Xstrahl Ltd and Johns Hopkins University.
OSTI ID:
22679075
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Systematic study of target localization for bioluminescence tomography guided radiation therapy
Journal Article · Sun May 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22620910

Bioluminescence Tomography–Guided Radiation Therapy for Preclinical Research
Journal Article · Fri Apr 01 00:00:00 EDT 2016 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22645157

SU-C-303-04: Evaluation of On- and Off-Line Bioluminescence Tomography System for Focal Irradiation Guidance
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22486563