Title: TU-H-CAMPUS-IeP2-03: Development of 3D Printed Coronary Phantoms for In-Vitro CT-FFR Validation Using Data from 320- Detector Row Coronary CT Angiography

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4957681· OSTI ID:22654058
; ; ; ;  [1];  [2];  [3];  [4]
  1. University at Buffalo, Buffalo, NY (United States)
  2. Toshiba America Medical Systems, Inc, Tustin, CA (United States)
  3. Brigham and Women’s Hospital, Boston, MA (United States)
  4. University of Ottawa, Ottawa, ON (Canada)

Purpose: To validate Computed Tomography Fractional Flow Reserve (CT-FFR) measurements with accurate 3D printed coronary phantoms. Methods: DICOM data from four phases in two patients imaged with a standard 320 × 0.5mm coronary CT acquisition (70–80% cardiac cycle) underwent semi-automated segmentation using a research workstation. Both patients had a >50% stenosis from the clinical image interpretation. Each volume was saved as a Stereo Lithographic (STL) file with 250 micron resolution. The 3D geometries were qualitatively assessed; the best of the four phases was 3D printed using a Stratasys Eden260V printer in Tango+, a rubber-like material that roughly emulates mechanical properties of human vasculature. We connected the model to a programmable pump and measured the pressure drop using pressure sensors embedded proximal and distal to the arterial stenosis. Next, the STL files used for the 3D printed models were uploaded in the ANSYS meshing tool (ICEM CFD 16.1). A standard meshing process was applied and the meshed geometry was directly imported in the ANSYS Fluent for Computational Flow Dynamics simulations. The CFD simulations were used to calculate the CT-FFR and compared to the bench top FFR measured in the 3D printed phantoms. Results: FFR-CT measurements and phantoms were completed in within an hour after the segmentation. Patient 1 had a 60% stenosis that resulted in a CT-FFR of 0.68. The second case had a 50% stenosis and a CT-FFR of 0.75. The average bench top FFR measurements were 0.72 and 0.80, respectively. Conclusion: This pilot investigation demonstrated the use of a bench-top coronary model for CT-FFR validation. The measurements and the CFD simulations agreed within 6%. Project supported by Support: Toshiba America Medical Systems Corp.and NIH grant R01-EB002873. Project supported by Toshiba America Medical Systems Corp.and partial support from NIH grant R01-EB002873.

OSTI ID:
22654058
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English