TU-H-CAMPUS-IeP1-02: Validation of a CT Monte Carlo Software
- Technische Hochschule Mittelhessen - University of Applied Sciences, Giessen (Germany)
Purpose: To validate the in-house developed CT Monte Carlo calculation tool GMctdospp against reference simulation sets provided by the AAPM in the new report 195. Methods: Deposited energy was calculated in four segments (test 1) and two 10 cm long cylinders (test 2) inside a CTDI phantom (following case #4 of the AAPM report 195). The x-ray point source of a given 120 kVp spectrum was collimated to a fan beam with two thicknesses (10 mm, 80 mm) for a static and a rotational setup. In addition, a given chest geometry was used to calculate deposited energy in several organs for a 0° static and a rotational beam (following case #5 of the AAPM report 195). The results of GMctdospp were compared against the particular mean value of the four quoted Monte Carlo codes (EGSnrc, Geant 4, MCNP and Penelope). Results: Calculated values showed no outliers in any of the cases. Differences between GMctdospp and the particular mean Results: Calculated values showed no outliers in any of the cases. Differences between GMctdospp and the particular mean value were always at similar magnitude compared to the quoted codes. For case #4 (CTDI phantom) the relative differences were within 1.5 %, on average 0.4 % and for case #5 (chest phantom) within 2.5 % and on average 0.85 %. Conclusion: The results confirmed an overall uncertainty of the Monte-Carlo calculation chain in GMctdospp being <2.5 %, for most cases even better. This can be considered small compared to other sources of uncertainties, e.g. virtual source and patient models. The photon transport implemented in GMctdospp inside a voxel-based patient geometry was successfully verified.
- OSTI ID:
- 22654045
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
GMctdospp: Description and validation of a CT dose calculation system
SU-F-207-01: Comparison of Beam Characteristics and Organ Dose From Four Commercial Multidetector Computed Tomography Scanners
SU-F-J-14: Kilovoltage Cone-Beam CT Dose Estimation of Varian On-Board Imager Using GMctdospp Monte Carlo Framework
Journal Article
·
Wed Jul 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22581381
SU-F-207-01: Comparison of Beam Characteristics and Organ Dose From Four Commercial Multidetector Computed Tomography Scanners
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22555231
SU-F-J-14: Kilovoltage Cone-Beam CT Dose Estimation of Varian On-Board Imager Using GMctdospp Monte Carlo Framework
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22632150