Title: TU-H-206-02: Novel Linearly-Filled Derenzo PET Phantom Design

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4957647· OSTI ID:22654039
; ; ; ; ; ;  [1];  [2]
  1. University of Wisconsin - Madison (United States)
  2. SOFIE Biosciences, Culver City, CA (United States)

Purpose: To design a linearly-filled Derenzo positron emission tomography (PET) phantom, eliminating the extraneous radioisotope volumes in a conventional reservoir-type design. This activity reduction combined with the elimination of bubbles in smaller phantom channels would significantly reduce personnel dose, radioisotope cost, and would improve image quality by reducing out-of-slice activity scatter. Methods: A computer-aided design (CAD) was created of a modular Derenzo phantom consisting of three phantom layers with gaskets between the layers. The central piece contains the active pattern volume and channels connecting adjacent rods in a serpentine pattern. The two end-pieces contained an inlet and an outlet for filling purposes. Phantom prototypes were 3D printed on a Viper Si2 stereolithography machine. The two gaskets were fabricated from silicon sheets using a PLS 6.75 laser cutter. Phantoms were held together by pass-through glass-filled nylon bolts and nuts. Phantoms were filled with {sup 52}Mn, {sup 64}Cu, {sup 74}Br, and {sup 124}I for testing, and were imaged on a Siemens Inveon MicroPET scanner. Results: Four phantom prototypes were constructed using male Leur Lock fittings for inlet/outlet ports. 3D printed layers were sanded to ensure proper coupling to the silicon gaskets. The filling volume for each prototype was approximately 2.4 mL. The filling process was found to be rapid, leak-tight, and with minimal back-pressure. PET images were reconstructed by OSEM3D, and axial slices along the phantom pattern length were averaged to provide final images. Image distortion was isotope dependent with {sup 52}Mn and {sup 64}Cu having the least distortion and {sup 124}I having the most distortion. Conclusion: These results indicate that the linearlyfilled Derenzo design improves on conventional reservoir-type designs by eliminating potential bubbles in small channels and by reducing activity level, radioisotope volume, radioisotope cost, personnel dose, filling time, and out-of-slice activity scatter The method described in this abstract has been filed as a patent application to the US Patent and Trade Office by the Wisconsin Alumni Research Foundation (WARF).

OSTI ID:
22654039
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English