TU-AB-202-02: Deformable Image Registration Accuracy Between External Beam Radiotherapy and HDR Brachytherapy CT Images for Cervical Cancer Using a 3D-Printed Deformable Pelvis Phantom
- Tohoku University Graduate School of Medicine, Sendai, Miyagi (Japan)
- Cybermedia Center, Osaka University, Toyonaka, Osaka (Japan)
- Tohoku University Hospital, Sendai, Miyagi (Japan)
Purpose: Accurate deformable image registration (DIR) between external beam radiotherapy (EBRT) and HDR brachytherapy (BT) CT images in cervical cancer is challenging. DSC has been evaluated only on the basis of the consistency of the structure, and its use does not guarantee an anatomically reasonable deformation. We evaluate the DIR accuracy for cervical cancer with DSC and anatomical landmarks using a 3D-printed pelvis phantom. Methods: A 3D-printed, deformable female pelvis phantom was created on the basis of the patient’s CT image. Urethane and silicon were used as materials for creating the uterus and bladder, respectively, in the phantom. We performed DIR in two cases: case-A with a full bladder (170 ml) in both the EBRT and BT images and case-B with a full bladder in the BT image and a half bladder (100 ml) in the EBRT image. DIR was evaluated using DSCs and 70 uterus and bladder landmarks. A Hybrid intensity and structure DIR algorithm with two settings (RayStation) was used. Results: In the case-A, DSCs of the intensity-based DIR were 0.93 and 0.85 for the bladder and uterus, respectively, whereas those of hybrid-DIR were 0.98 and 0.96, respectively. The mean landmark error values of intensity-based DIR were 0.73±0.29 and 1.70±0.19 cm for the bladder and uterus, respectively, whereas those of Hybrid-DIR were 0.43±0.33 and 1.23±0.25 cm, respectively. In both cases, the Hybrid-DIR accuracy was better than the intensity-based DIR accuracy for both evaluation methods. However, for several bladder landmarks, the Hybrid-DIR landmark errors were larger than the corresponding intensity-based DIR errors (e.g., 2.26 vs 1.25 cm). Conclusion: Our results demonstrate that Hybrid-DIR can perform with a better accuracy than the intensity-based DIR for both DSC and landmark errors; however, Hybrid-DIR shows a larger landmark error for some landmarks because the technique focuses on both the structure and intensity.
- OSTI ID:
- 22653955
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-F-J-88: Comparison of Two Deformable Image Registration Algorithms for CT-To-CT Contour Propagation
Improving anatomical mapping of complexly deformed anatomy for external beam radiotherapy and brachytherapy dose accumulation in cervical cancer
SU-F-J-97: A Joint Registration and Segmentation Approach for Large Bladder Deformations in Adaptive Radiotherapy
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22632213
Improving anatomical mapping of complexly deformed anatomy for external beam radiotherapy and brachytherapy dose accumulation in cervical cancer
Journal Article
·
Wed Jan 14 23:00:00 EST 2015
· Medical Physics
·
OSTI ID:22413376
SU-F-J-97: A Joint Registration and Segmentation Approach for Large Bladder Deformations in Adaptive Radiotherapy
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22634706