MO-FG-202-04: Gantry-Resolved Linac QA for VMAT: A Comprehensive and Efficient System Using An Electronic Portal Imaging Device
- Central Coast Cancer Centre, Gosford, NSW (Australia)
- University of Newcastle, Newcastle, NSW (Australia)
- University of Sydney, Camperdown, NSW (Australia)
Purpose: To automate gantry-resolved linear accelerator (linac) quality assurance (QA) for volumetric modulated arc therapy (VMAT) using an electronic portal imaging device (EPID). Methods: A QA system for VMAT was developed that uses an EPID, frame-grabber assembly and in-house developed image processing software. The system relies solely on the analysis of EPID image frames acquired without the presence of a phantom. Images were acquired at 8.41 frames per second using a frame grabber and ancillary acquisition computer. Each image frame was tagged with a gantry angle from the linac’s on-board gantry angle encoder. Arc-dynamic QA plans were designed to assess the performance of each individual linac component during VMAT. By analysing each image frame acquired during the QA deliveries the following eight machine performance characteristics were measured as a function of gantry angle: MLC positional accuracy, MLC speed constancy, MLC acceleration constancy, MLC-gantry synchronisation, beam profile constancy, dose rate constancy, gantry speed constancy, dose-gantry angle synchronisation and mechanical sag. All tests were performed on a Varian iX linear accelerator equipped with a 120 leaf Millennium MLC and an aS1000 EPID (Varian Medical Systems, Palo Alto, CA, USA). Results: Machine performance parameters were measured as a function of gantry angle using EPID imaging and compared to machine log files and the treatment plan. Data acquisition is currently underway at 3 centres, incorporating 7 treatment units, at 2 weekly measurement intervals. Conclusion: The proposed system can be applied for streamlined linac QA and commissioning for VMAT. The set of test plans developed can be used to assess the performance of each individual components of the treatment machine during VMAT deliveries as a function of gantry angle. The methodology does not require the setup of any additional phantom or measurement equipment and the analysis is fully automated to allow for regular routine testing.
- OSTI ID:
- 22653875
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Using an EPID for patient-specific VMAT quality assurance
Frame average optimization of cine-mode EPID images used for routine clinical in vivo patient dose verification of VMAT deliveries
Gantry-angle resolved VMAT pretreatment verification using EPID image prediction
Journal Article
·
Tue Mar 15 00:00:00 EDT 2011
· Medical Physics
·
OSTI ID:22096925
Frame average optimization of cine-mode EPID images used for routine clinical in vivo patient dose verification of VMAT deliveries
Journal Article
·
Thu Jan 14 23:00:00 EST 2016
· Medical Physics
·
OSTI ID:22579810
Gantry-angle resolved VMAT pretreatment verification using EPID image prediction
Journal Article
·
Thu Aug 15 00:00:00 EDT 2013
· Medical Physics
·
OSTI ID:22220512