Title: SU-G-TeP4-07: Automatic EPID-Based 2D Measurement of MLC Leaf Offset as a Quality Control Tool

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4957132· OSTI ID:22649470
;  [1];  [2];  [3];  [4];  [5];  [6];  [7];  [8]
  1. The University of Michigan, Ann Arbor, MI (United States)
  2. University of Michigan, Ann Arbor, MI (United States)
  3. University of California, San Diego, La Jolla, CA (United States)
  4. Calvary Mater Hospital Newcastle, Warratah, NSW (Australia)
  5. North Sydney Cancer Center, Sydney (Australia)
  6. University of Chicago, Chicago, IL (United States)
  7. University Alabama Birmingham, Birmingham, AL (United States)
  8. Calvary Mater Newcastle, Newcastle (Australia)

Purpose: The MLC dosimetric leaf gap (DLG) and transmission are measured parameters which impact the dosimetric accuracy of IMRT and VMAT plans. This investigation aims to develop an efficient and accurate routine constancy check of the physical DLG in two dimensions. Methods: The manufacturer’s recommended DLG measurement method was modified by using 5 fields instead of 11 and by utilizing the Electronic Portal Imaging Device (EPID). Validations were accomplished using an ion chamber (IC) in solid water and a 2D IC array. EPID data was collected for 6 months on multiple TrueBeam linacs using both Millennium and HD MLCs at 5 different clinics in an international consortium. Matlab code was written to automatically analyze the images and calculate the 2D results. Sensitivity was investigated by introducing deliberate leaf position errors. MLC calibration and initialization history was recorded to allow quantification of their impact. Results were analyzed using statistical process control (SPC). Results: The EPID method took approximately 5 minutes. Due to detector response, the EPID measured DLG and transmission differed from the IC values but were reproducible and consistent with changes measured using the ICs. For the Millennium MLC, the EPID measured DLG and transmission were both consistently lower than IC results. The EPID method was implemented as leaf offset and transmission constancy tests (LOC and TC). Based on 6 months of measurements, the initial leaf-specific action thresholds for changes from baseline were set to 0.1 mm. Upper and lower control limits for variation were developed for each machine. Conclusion: Leaf offset and transmission constancy tests were implemented on Varian HD and Millennium MLCs using an EPID and found to be efficient and accurate. The test is effective for monitoring MLC performance using dynamic delivery and performing process control on the DLG in 2D, thus enhancing dosimetric accuracy. This work was supported by a grant from Varian Medical Systems.

OSTI ID:
22649470
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Daily QA of linear accelerators using only EPID and OBI
Journal Article · Thu Oct 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22649470

Verification of dosimetric accuracy on the TrueBeam STx: Rounded leaf effect of the high definition MLC
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22649470

SU-E-T-01: 2-D Characterization of DLG Among All MLC Leaf Pairs
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22649470