Title: SU-G-TeP2-12: IROCHouston and MDAPL SRS Anthropomorphic Phantom Results

Abstract

Purpose: To report the results of SRS phantom irradiations Methods: Anthropomorphic SRS head phantoms were sent to institutions participating in NCI sponsored SRS clinical trials and institutions interested in verifying SRS treatment delivery. The phantom shell was purchased from Phantom Laboratory and altered to house dosimetry and imaging inserts. The imaging insert has 1.9 cm diameter spherical target. The dosimetry insert holds two TLD capsules and radiochromic film in the coronal and sagittal planes through the center of the target. Institutions were asked to image, plan and treat the phantom as they would an SRS patient. GammaKnife, CyberKnife and c-arm accelerator institutions were asked to cover the target with 15 Gy, 20 Gy and 25 Gy, respectively. Following these guidelines and typical planning protocols for these three types of machines gives roughly 30 Gy to the center of the target for all units. Submission of the DICOM digital data set was required for analysis. Criteria of 5% for TLD results and 85% of pixels passing 5%/3mm gamma analysis were applied beginning in 2013. Results: The phantom was analyzed 269 times between the beginning of 2013 to present. The pass rate is 81%. Nineteen of the irradiation results failed only themore » TLD criteria, 19 failed only the film criteria and 12 failed both. Irradiations included 32 CyberKnife 23 GammaKnife, 3 TomoTherapy and 211 c-arm units. Planning systems included Eclipse, Ergo, GammaPlan, Hi-Art, iPlan, Monaco, MultiPlan, Pinnacle, RayStation, XiO and XKnife. Irradiations that were not accompanied with DICOM data were not included in this analysis. Conclusion: The phantom is a valuable end-to-end test used to independently verify the accuracy of SRS treatment delivery. This investigation was supported by IROC grant CA180803 awarded by the NCI.« less

Authors:
; ; ; ; ;  [1]
  1. UT MD Anderson Cancer Center, Houston, TX (United States)
Publication Date:
OSTI Identifier:
22649392
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; CLINICAL TRIALS; COMPUTERIZED TOMOGRAPHY; CT-GUIDED RADIOTHERAPY; IMAGES; IRRADIATION; PHANTOMS; PLANNING; SPHERICAL CONFIGURATION

Citation Formats

Molineu, A, Kry, S, Alvarez, P, Hernandez, N, Nguyen, T, and Followill, D. SU-G-TeP2-12: IROCHouston and MDAPL SRS Anthropomorphic Phantom Results. United States: N. p., 2016. Web. doi:10.1118/1.4957047.
Molineu, A, Kry, S, Alvarez, P, Hernandez, N, Nguyen, T, & Followill, D. SU-G-TeP2-12: IROCHouston and MDAPL SRS Anthropomorphic Phantom Results. United States. doi:10.1118/1.4957047.
Molineu, A, Kry, S, Alvarez, P, Hernandez, N, Nguyen, T, and Followill, D. 2016. "SU-G-TeP2-12: IROCHouston and MDAPL SRS Anthropomorphic Phantom Results". United States. doi:10.1118/1.4957047.
@article{osti_22649392,
title = {SU-G-TeP2-12: IROCHouston and MDAPL SRS Anthropomorphic Phantom Results},
author = {Molineu, A and Kry, S and Alvarez, P and Hernandez, N and Nguyen, T and Followill, D},
abstractNote = {Purpose: To report the results of SRS phantom irradiations Methods: Anthropomorphic SRS head phantoms were sent to institutions participating in NCI sponsored SRS clinical trials and institutions interested in verifying SRS treatment delivery. The phantom shell was purchased from Phantom Laboratory and altered to house dosimetry and imaging inserts. The imaging insert has 1.9 cm diameter spherical target. The dosimetry insert holds two TLD capsules and radiochromic film in the coronal and sagittal planes through the center of the target. Institutions were asked to image, plan and treat the phantom as they would an SRS patient. GammaKnife, CyberKnife and c-arm accelerator institutions were asked to cover the target with 15 Gy, 20 Gy and 25 Gy, respectively. Following these guidelines and typical planning protocols for these three types of machines gives roughly 30 Gy to the center of the target for all units. Submission of the DICOM digital data set was required for analysis. Criteria of 5% for TLD results and 85% of pixels passing 5%/3mm gamma analysis were applied beginning in 2013. Results: The phantom was analyzed 269 times between the beginning of 2013 to present. The pass rate is 81%. Nineteen of the irradiation results failed only the TLD criteria, 19 failed only the film criteria and 12 failed both. Irradiations included 32 CyberKnife 23 GammaKnife, 3 TomoTherapy and 211 c-arm units. Planning systems included Eclipse, Ergo, GammaPlan, Hi-Art, iPlan, Monaco, MultiPlan, Pinnacle, RayStation, XiO and XKnife. Irradiations that were not accompanied with DICOM data were not included in this analysis. Conclusion: The phantom is a valuable end-to-end test used to independently verify the accuracy of SRS treatment delivery. This investigation was supported by IROC grant CA180803 awarded by the NCI.},
doi = {10.1118/1.4957047},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: To analyze the results from an anthropomorphic spine phantom used for credentialing institutions for National Cancer Institute (NCI) sponsored clinical trial. Methods: An anthropomorphic phantom that contains left and right lungs, a heart, an esophagus, spinal cord, bony material and a PTV was sent to institutions wishing to be credentialed for NCI trials. The PTV holds 4 TLD and radiochromic film in the axial and sagittal planes. The heart holds one TLD. Institutions created IMRT plans to cover ≥90% of the PTV with 6 Gy and limit the cord dose to <0.35cc receiving 3.75 Gy and <1.2cc receiving 2.63more » Gy. They were instructed to treat the phantom as they would a patient, including making plan specific IMRT/SBRT QA measurements before treatment. The TLD results in the PTV were required to be within ±7% of the plan dose. A gamma calculation was performed using the film results and the submitted DICOM plan. ≥85% of the analyzed region was required to pass a 5%/3 mm criteria. Results: 176 institutions irradiated the spine phantom for a total of 255 results. The pass rate was 73% (187 irradiations) overall. 44 irradiations failed only the gamma criteria, 2 failed only the dose criteria and 22 failed both. The most used planning systems were Eclipse (116) and Pinnacle (52) and they had pass rates of 76% and 71%, respectively. The AAA algorithm had a pass rate of 77% while superposition type algorithms had a 71% pass rate. The average TLD measurement to institution calculation ratio was 0.99 (0.04 std dev.). The average percent pixels passing the gamma criteria for films was 89% (12% std dev.) Conclusion: Results show that this phantom is an important part of credentialing and that we have room for improvement in IMRT/SBRT spine treatments. This work was supported by PHS CA180803 and CA037422 awarded by NCI, DHHS.« less
  • Purpose: To measure radiation absorbed dose and its distribution in an anthropomorphic head phantom under clinically representative scan conditions in three widely used computed tomography (CT) scanners, and to relate those dose values to metrics such as high-contrast resolution, noise, and contrast-to-noise ratio (CNR) in the American College of Radiology CT accreditation phantom.Methods: By inserting optically stimulated luminescence dosimeters (OSLDs) in the head of an anthropomorphic phantom specially developed for CT dosimetry (University of Florida, Gainesville), we measured dose with three commonly used scanners (GE Discovery CT750 HD, Siemens Definition, Philips Brilliance 64) at two different clinical sites (Walter Reedmore » National Military Medical Center, National Institutes of Health). The scanners were set to operate with the same data-acquisition and image-reconstruction protocols as used clinically for typical head scans, respective of the practices of each facility for each scanner. We also analyzed images of the ACR CT accreditation phantom with the corresponding protocols. While the Siemens Definition and the Philips Brilliance protocols utilized only conventional, filtered back-projection (FBP) image-reconstruction methods, the GE Discovery also employed its particular version of an adaptive statistical iterative reconstruction (ASIR) algorithm that can be blended in desired proportions with the FBP algorithm. We did an objective image-metrics analysis evaluating the modulation transfer function (MTF), noise power spectrum (NPS), and CNR for images reconstructed with FBP. For images reconstructed with ASIR, we only analyzed the CNR, since MTF and NPS results are expected to depend on the object for iterative reconstruction algorithms.Results: The OSLD measurements showed that the Siemens Definition and the Philips Brilliance scanners (located at two different clinical facilities) yield average absorbed doses in tissue of 42.6 and 43.1 mGy, respectively. The GE Discovery delivers about the same amount of dose (43.7 mGy) when run under similar operating and image-reconstruction conditions, i.e., without tube current modulation and ASIR. The image-metrics analysis likewise showed that the MTF, NPS, and CNR associated with the reconstructed images are mutually comparable when the three scanners are run with similar settings, and differences can be attributed to different edge-enhancement properties of the applied reconstruction filters. Moreover, when the GE scanner was operated with the facility's scanner settings for routine head exams, which apply 50% ASIR and use only approximately half of the 100%-FBP dose, the CNR of the images showed no significant change. Even though the CNR alone is not sufficient to characterize the image quality and justify any dose reduction claims, it can be useful as a constancy test metric.Conclusions: This work presents a straightforward method to connect direct measurements of CT dose with objective image metrics such as high-contrast resolution, noise, and CNR. It demonstrates that OSLD measurements in an anthropomorphic head phantom allow a realistic and locally precise estimation of magnitude and spatial distribution of dose in tissue delivered during a typical CT head scan. Additional objective analysis of the images of the ACR accreditation phantom can be used to relate the measured doses to high contrast resolution, noise, and CNR.« less
  • Concern about the amount of radiation received during scoliosis evaluation and treatment led us to measure radiation exposure in an anthropomorphic phantom to determine the increased risk of breast cancer in young women with scoliosis. Assuming that 22 radiographic examinations were performed over the course of scoliosis treatment, the increased relative risk of breast cancer was determined to be 0.22% in these patients.
  • The purpose of this work was to evaluate lesion detectability with and without nonuniform attenuation compensation (AC) in myocardial perfusion SPECT imaging in women using an anthropomorphic phantom and receiver operating characteristics (ROC) methodology. Breast attenuation causes artifacts in reconstructed images and may increase the difficulty of diagnosis of myocardial perfusion imaging in women. The null hypothesis tested using the ROC study was that nonuniform AC does not change the lesion detectability in myocardial perfusion SPECT imaging in women. The authors used a filtered backprojection (FBP) reconstruction algorithm and Chang`s single iteration method for AC. In conclusion, with the proposedmore » myocardial defect model nuclear medicine physicians demonstrated no significant difference for the detection of the anterior wall defect; however, a greater accuracy for the detection of the inferior wall defect was observed without nonuniform AC than with it. Medical physicists did not demonstrate any statistically significant difference in defect detection accuracy with or without nonuniform AC in the female phantom.« less