Title: SU-G-JeP2-06: Dosimetric and Workflow Evaluation of First Commercial Synthetic CT Software for Clinical Use in Pelvis

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4957026· OSTI ID:22649372
; ; ; ; ; ;  [1]
  1. Memorial Sloan Kettering Cancer Center, New York, NY (United States)

Purpose: evaluate a commercial synthetic CT (syn-CT) software for use in prostate radiotherapy Methods: Twenty prostate patients underwent CT and MR simulation scans in treatment position on a 3T Philips scanner. The MR protocol consisted of a T2w turbo spin-echo for soft tissue contrast, a 2D balanced-fast field echo (b-FFE) for fiducial identification, a dual-echo 3D FFE B0 map for distortion analysis and a 3D mDIXON FFE sequence to generate syn-CT. Two echoes are acquired during mDIXON scan, allowing water, fat, and in-phase images to be derived using the frequency shift of the fat and water protons. Tissues were classified as: air, adipose, water, trabecular/spongy bone and compact/cortical bone and assigned specific bulk HU values. Bone structures are segmented based on a pelvis bone atlas. Accuracy of syn-CT for patient treatment planning was analyzed by transferring the original plan and structures from the CT to syn-CT via rigid registration and recalculating dose. In addition, new IMRT plans were generated on the syn-CT using structures contoured on MR and transferred to the syn-CT. Accuracy of fiducial-based localization at the treatment machine performed using syn-CT or DRRs generated from syn-CT was assessed by comparing to orthogonal kV radiographs or CBCT. Results: Dosimetric comparison between CT and syn-CT was within 0.5% for all structures. The de-novo optimized plans generated on the syn-CT met our institutional clinical objectives for target and normal structures. Patient-induced susceptibility distortion based on B0 maps was within 1mm and 0.4 mm in the body and prostate. The rectal and bladder outlines on the syn-CT were deemed sufficient for assessing rectal and bladder filling on the CBCT at the time of treatment. CBCT localization showed a median error of < ±1 mm in LR, AP and SI direction. Conclusion: MRI derived syn-CT can be used clinically in MR-alone planning and treatment process for prostate. Drs. Deasy, Hunt and Tyagi have Master research agreement with Philips healthcare.

OSTI ID:
22649372
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English