Title: SU-F-T-536: Contra-Lateral Breast Study for Prone Versus Supine Patients

Abstract

Purpose: There are several advantages to utilizing the prone technique for intact breast cancer patients. However, as the topography changes, accompanied by the influence of a supporting breast board and patient treatment couch, the question that arises is to whether there is a concern for contralateral breast dose for intact breast cancer patients being treated with this technique. Methods: An anthropomorphic phantom with breast mounds to duplicate intact breast cancer treatment was planned in prone and supine position. Two tangential beams were executed in the similar manner for as the radiotherapy planning system. For the prone setup, a breast dense foam board was used to support the phantom. A grid of 24 OSL nanodots was placed at 6cm, 4cm, and 2cm apart from the medial border for both prone and supine setups. The phantom was set up using megavoltage imaging and treated as per plan. Additional, a similar study was performed on a patient treated in prone position. Results: Overall, the contralateral breast dose was generally higher for prone setups at all locations especially when close to the medial border. The average mean dose was found to be 1.8%, 2.5% of the prescribed dose for supine respectively prone position. Themore » average of the standard deviation is 1.04%, 1.38% for supine respectively prone position. As for patient treated in prone position average mean dose was found to be 1.165% of the prescribed dose and average of the standard deviation is 9.456%. Conclusion: There is minimal influence of scatter from the breast board. It appears that the volatility of the setup could lead to higher doses than expected from the planning system to the contralateral breast when the patient is in the prone position.« less

Authors:
; ;  [1]
  1. Northwell Health, Lake Success, NY (United States)
Publication Date:
OSTI Identifier:
22649117
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; BIOMEDICAL RADIOGRAPHY; MAMMARY GLANDS; NEOPLASMS; PATIENTS; PHANTOMS; PLANNING; QUANTUM DOTS; RADIATION DOSES

Citation Formats

Marrero, M, Joseph, K, and Klein, E. SU-F-T-536: Contra-Lateral Breast Study for Prone Versus Supine Patients. United States: N. p., 2016. Web. doi:10.1118/1.4956721.
Marrero, M, Joseph, K, & Klein, E. SU-F-T-536: Contra-Lateral Breast Study for Prone Versus Supine Patients. United States. doi:10.1118/1.4956721.
Marrero, M, Joseph, K, and Klein, E. 2016. "SU-F-T-536: Contra-Lateral Breast Study for Prone Versus Supine Patients". United States. doi:10.1118/1.4956721.
@article{osti_22649117,
title = {SU-F-T-536: Contra-Lateral Breast Study for Prone Versus Supine Patients},
author = {Marrero, M and Joseph, K and Klein, E},
abstractNote = {Purpose: There are several advantages to utilizing the prone technique for intact breast cancer patients. However, as the topography changes, accompanied by the influence of a supporting breast board and patient treatment couch, the question that arises is to whether there is a concern for contralateral breast dose for intact breast cancer patients being treated with this technique. Methods: An anthropomorphic phantom with breast mounds to duplicate intact breast cancer treatment was planned in prone and supine position. Two tangential beams were executed in the similar manner for as the radiotherapy planning system. For the prone setup, a breast dense foam board was used to support the phantom. A grid of 24 OSL nanodots was placed at 6cm, 4cm, and 2cm apart from the medial border for both prone and supine setups. The phantom was set up using megavoltage imaging and treated as per plan. Additional, a similar study was performed on a patient treated in prone position. Results: Overall, the contralateral breast dose was generally higher for prone setups at all locations especially when close to the medial border. The average mean dose was found to be 1.8%, 2.5% of the prescribed dose for supine respectively prone position. The average of the standard deviation is 1.04%, 1.38% for supine respectively prone position. As for patient treated in prone position average mean dose was found to be 1.165% of the prescribed dose and average of the standard deviation is 9.456%. Conclusion: There is minimal influence of scatter from the breast board. It appears that the volatility of the setup could lead to higher doses than expected from the planning system to the contralateral breast when the patient is in the prone position.},
doi = {10.1118/1.4956721},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: Damage to heart and lung from breast radiotherapy is associated with increased cardiovascular mortality and lung cancer development. We conducted a prospective study to evaluate which position is best to spare lung and heart from radiotherapy exposure. Methods and Materials: One hundred consecutive Stage 0-IIA breast cancer patients consented to participate in a research trial that required two computed tomography simulation scans for planning both supine and prone positions. The optimal position was defined as that which best covered the contoured breast and tumor bed while it minimized critical organ irradiation, as quantified by the in-field heart and lungmore » volume. The trial was designed to plan the first 100 patients in each position to study correlations between in-field volumes of organs at risk and dose. Results: Fifty-three left and 47 right breast cancer patients were consecutively accrued to the trial. In all patients, the prone position was optimal for sparing lung volume compared to the supine setup (mean lung volume reduction was 93.5 cc for right and 103.6 cc for left breast cancer patients). In 46/53 (87%) left breast cancer patients best treated prone, in-field heart volume was reduced by a mean of 12 cc and by 1.8 cc for the other 7/53 (13%) patients best treated supine. As predicted, supine-prone differences in in-field volume and mean dose of heart and lung were highly correlated (Spearman's correlation coefficient for left breast cancer patients was 0.90 for heart and 0.94 for lung and 0.92 for right breast cancer patients for lung). Conclusions: Prone setup reduced the amount of irradiated lung in all patients and reduced the amount of heart volume irradiated in 87% of left breast cancer patients. In-field organ volume is a valid surrogate for predicting dose; the trial continued to the planned target of 400.« less
  • Purpose: To study breast radiotherapy in the prone vs. supine positions through dosimetry and clinical implementation. Methods and Materials: Conformal radiotherapy plans in 61 patients requiring only breast irradiation were developed for both the prone and supine positions. After evaluation of the of the first 20 plan pairs, the patients were irradiated in the prone or supine position in a randomized fashion. These cases were analyzed for repositioning accuracy and skin reactions related to treatment position and patient characteristics. Results: The planning target volume covered with 47.5-53.5 Gy in the prone vs. the supine position was 85.1% {+-} 4.2% vs.more » 89.2 {+-} 2.2%, respectively (p < 0.0001). Radiation exposure of the ipsilateral lung, expressed in terms of the mean lung dose and the V{sub 20Gy}, was dramatically lower in the prone vs. supine position (p < 0.0001), but the doses to the heart did not differ. There was no difference in the need to correct positioning during radiotherapy, but the extent of displacement was significantly higher in the prone vs. supine position (p = 0.021). The repositioning accuracy in the prone position exhibited an improvement over time and did not depend on any patient-related parameters. Significantly more radiodermatitis of Grade 1-2 developed following prone vs. supine irradiation (p = 0.025). Conclusions: Conformal breast radiotherapy is feasible in the prone position. Its primary advantage is the substantially lower radiation dose to the ipsilateral lung. The higher dose inhomogeneity and increased rate of Grade 1-2 skin toxicity, however, may be of concern.« less
  • Purpose: To compare dosimetrically four different techniques of accelerated partial breast irradiation (APBI) in the same patient. Methods and Materials: Thirteen post-lumpectomy interstitial brachytherapy (IB) patients underwent imaging with preimplant computed tomography (CT) in the prone and supine position. These CT scans were then used to generate three-dimensional conformal radiotherapy (3D-CRT) and prone and supine helical tomotherapy (PT and ST, respectively) APBI plans and compared with the treated IB plans. Dose-volume histogram analysis and the mean dose (NTD{sub mean}) values were compared. Results: Planning target volume coverage was excellent for all methods. Statistical significance was considered to be a pmore » value <0.05. The mean V100 was significantly lower for IB (12% vs. 15% for PT, 18% for ST, and 26% for 3D-CRT). A greater significant differential was seen when comparing V50 with mean values of 24%, 43%, 47%, and 52% for IB, PT, ST, and 3D-CRT, respectively. The IB and PT were similar and delivered an average lung NTD{sub mean} dose of 1.3 Gy{sub 3} and 1.2 Gy{sub 3}, respectively. Both of these methods were statistically significantly lower than the supine external beam techniques. Overall, all four methods yielded similar low doses to the heart. Conclusions: The use of IB and PT resulted in greater normal tissue sparing (especially ipsilateral breast and lung) than the use of supine external beam techniques of 3D-CRT or ST. However, the choice of APBI technique must be tailored to the patient's anatomy, lumpectomy cavity location, and overall treatment goals.« less
  • Purpose: Previous studies have demonstrated how imaging of the breast with patients lying prone using a supportive positioning device markedly facilitates longitudinal and/or multimodal image registration. In this contribution, the authors’ primary objective was to determine if there are differences in the standardized uptake value (SUV) derived from [{sup 18}F]fluorodeoxyglucose (18F-FDG) positron emission tomography (PET) in breast tumors imaged in the standard supine position and in the prone position using a specialized positioning device. Methods: A custom positioning device was constructed to allow for breast scanning in the prone position. Rigid and nonrigid phantom studies evaluated differences in prone andmore » supine PET. Clinical studies comprised 18F-FDG-PET of 34 patients with locally advanced breast cancer imaged in the prone position (with the custom support) followed by imaging in the supine position (without the support). Mean and maximum values (SUV{sub peak} and SUV{sub max}, respectively) were obtained from tumor regions-of-interest for both positions. Prone and supine SUV were linearly corrected to account for the differences in 18F-FDG uptake time. Correlation, Bland–Altman, and nonparametric analyses were performed on uptake time-corrected and uncorrected data. Results: SUV from the rigid PET breast phantom imaged in the prone position with the support device was 1.9% lower than without the support device. In the nonrigid PET breast phantom, prone SUV with the support device was 5.0% lower than supine SUV without the support device. In patients, the median (range) difference in uptake time between prone and supine scans was 16.4 min (13.4–30.9 min), which was significantly—but not completely—reduced by the linear correction method. SUV{sub peak} and SUV{sub max} from prone versus supine scans were highly correlated, with concordance correlation coefficients of 0.91 and 0.90, respectively. Prone SUV{sub peak} and SUV{sub max} were significantly lower than supine in both original and uptake time-adjusted data across a range of index times (P < < 0.0001, Wilcoxon signed rank test). Before correcting for uptake time differences, Bland–Altman analyses revealed proportional bias between prone and supine measurements (SUV{sub peak} and SUV{sub max}) that increased with higher levels of FDG uptake. After uptake time correction, this bias was significantly reduced (P < 0.01). Significant prone-supine differences, with regard to the spatial distribution of lesions relative to isocenter, were observed between the two scan positions, but this was poorly correlated with the residual (uptake time-corrected) prone-supine SUV{sub peak} difference (P = 0.78). Conclusions: Quantitative 18F-FDG-PET/CT of the breast in the prone position is not deleteriously affected by the support device but yields SUV that is consistently lower than those obtained in the standard supine position. SUV differences between scans arising from FDG uptake time differences can be substantially reduced, but not removed entirely, with the current correction method. SUV from the two scan orientations is quantitatively different and should not be assumed equivalent or interchangeable within the same subject. These findings have clinical relevance in that they underscore the importance of patient positioning while scanning as a clinical variable that must be accounted for with longitudinal PET measurement, for example, in the assessment of treatment response.« less
  • Purpose: The prone treatment position has been used to reduce ipsilateral lung and heart dose in left breast radiation. We conducted a retrospective study to evaluate the difference in the dosimetry between prone and supine treatment positions. Methods: Eight left breast cancer patients were simulated in both the supine and prone positions as a pretreatment evaluation for the optimal treatment position. Treatment plans were created for all patients in both the supine and prone positions using a field in field three dimensional planning technique. Prescribed dose was 45 Gy delivered by two tangential photon fields. Irradiated volume (IV) was evaluatedmore » by V50, V100, and dose to lung and heart by V5, V10, V20, and the mean dose were evaluated. Results: All dosimetry metrics for both the supine and prone plans met our internal normal structure guidelines which are based on Quantec data. The average IVs (50% and 100%) were 2223cc and 1361cc prone, 2315cc and 1315cc supine. The average ipsilateral lung Mean dose (0.83Gy prone vs 5.8Gy supine), V5 (1.6% prone vs 20.9% supine), V10 (0.78% prone vs 15% supine) and V20 (0.36% prone vs 11% supine) were significantly lower in prone position. Heart Mean dose (1.4Gy prone vs 2.9Gy supine), V10 (1.4% prone vs 5.0% supine) and V20 (0.4% prone vs 3.5% supine) were found improved for all patients except one where the mean dose was the same and all other values were improved. Conclusion: The prone position offer preferable dosimetry for all patients planned in our study. These patients were chosen based on the physician’s belief that they would benefit from prone treatment either because they had large pendulous breasts or due to the amount of heart seen in the field on CT simulation.« less