Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

SU-F-T-465: Two Years of Radiotherapy Treatments Analyzed Through MLC Log Files

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956650· OSTI ID:22649056
 [1]; ;  [2];  [3]
  1. University of Texas HSC SA, New Braunfles, TX (United States)
  2. University of Texas HSC SA, San Antonio, TX (United States)
  3. Cancer Therapy and Research Center, San Antonio, TX (United States)

Purpose: To present treatment statistics of a Varian Novalis Tx using more than 90,000 Varian Dynalog files collected over the past 2 years. Methods: Varian Dynalog files are recorded for every patient treated on our Varian Novalis Tx. The files are collected and analyzed daily to check interfraction agreement of treatment deliveries. This is accomplished by creating fluence maps from the data contained in the Dynalog files. From the Dynalog files we have also compiled statistics for treatment delivery times, MLC errors, gantry errors and collimator errors. Results: The mean treatment time for VMAT patients was 153 ± 86 seconds while the mean treatment time for step & shoot was 256 ± 149 seconds. Patient’s treatment times showed a variation of 0.4% over there treatment course for VMAT and 0.5% for step & shoot. The average field sizes were 40 cm2 and 26 cm2 for VMAT and step & shoot respectively. VMAT beams contained and average overall leaf travel of 34.17 meters and step & shoot beams averaged less than half of that at 15.93 meters. When comparing planned and delivered fluence maps generated using the Dynalog files VMAT plans showed an average gamma passing percentage of 99.85 ± 0.47. Step & shoot plans showed an average gamma passing percentage of 97.04 ± 0.04. 5.3% of beams contained an MLC error greater than 1 mm and 2.4% had an error greater than 2mm. The mean gantry speed for VMAT plans was 1.01 degrees/s with a maximum of 6.5 degrees/s. Conclusion: Varian Dynalog files are useful for monitoring machine performance treatment parameters. The Dynalog files have shown that the performance of the Novalis Tx is consistent over the course of a patients treatment with only slight variations in patient treatment times and a low rate of MLC errors.

OSTI ID:
22649056
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-325: The New Evaluation Method of the VMAT Plan Delivery Using Varian DynaLog Files and Modulation Complexity Score (MCS)
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22355881

SU-E-T-576: Evaluation of Patient Specific VMAT QA Using Dynalog Files and Treatment Planning System
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369693

SU-E-T-252: On the Evaluation of Patient Specific IMRT QA Using EPID, Dynalog Files and Patient Anatomy
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22351075