Title: SU-F-T-368: Improved HPGe Detector Precise Efficiency Calibration with Monte Carlo Simulations and Radioactive Sources

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956553· OSTI ID:22648966
 [1]
  1. Vanderbilt University, Vanderbilt-Ingram Cancer Center, Nashville, TN 37232 (United States)

Purpose: To obtain an improved precise gamma efficiency calibration curve of HPGe (High Purity Germanium) detector with a new comprehensive approach. Methods: Both of radioactive sources and Monte Carlo simulation (CYLTRAN) are used to determine HPGe gamma efficiency for energy range of 0–8 MeV. The HPGe is a GMX coaxial 280 cm{sup 3} N-type 70% gamma detector. Using Momentum Achromat Recoil Spectrometer (MARS) at the K500 superconducting cyclotron of Texas A&M University, the radioactive nucleus {sup 24} Al was produced and separated. This nucleus has positron decays followed by gamma transitions up to 8 MeV from {sup 24} Mg excited states which is used to do HPGe efficiency calibration. Results: With {sup 24} Al gamma energy spectrum up to 8MeV, the efficiency for γ ray 7.07 MeV at 4.9 cm distance away from the radioactive source {sup 24} Al was obtained at a value of 0.194(4)%, by carefully considering various factors such as positron annihilation, peak summing effect, beta detector efficiency and internal conversion effect. The Monte Carlo simulation (CYLTRAN) gave a value of 0.189%, which was in agreement with the experimental measurements. Applying to different energy points, then a precise efficiency calibration curve of HPGe detector up to 7.07 MeV at 4.9 cm distance away from the source {sup 24} Al was obtained. Using the same data analysis procedure, the efficiency for the 7.07 MeV gamma ray at 15.1 cm from the source {sup 24} Al was obtained at a value of 0.0387(6)%. MC simulation got a similar value of 0.0395%. This discrepancy led us to assign an uncertainty of 3% to the efficiency at 15.1 cm up to 7.07 MeV. The MC calculations also reproduced the intensity of observed single-and double-escape peaks, providing that the effects of positron annihilation-in-flight were incorporated. Conclusion: The precision improved gamma efficiency calibration curve provides more accurate radiation detection and dose calculation for cancer radiotherapy treatment.

OSTI ID:
22648966
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English