SU-F-T-300: Impact of Electron Density Modeling of ArcCHECK Cylindricaldiode Array On 3DVH Patient Specific QA Software Tool Analysis

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956485· OSTI ID:22648908
; ;  [1]
  1. Sir HN RF Hospital, Mumbai, Maharashtra (India)

Purpose: 3DVH software is a patient specific quality assurance tool which estimates the 3D dose to the patient specific geometry with the help of Planned Dose Perturbation algorithm. The purpose of this study is to evaluate the impact of HU value of ArcCHECK phantom entered in Eclipse TPS on 3D dose & DVH QA analysis. Methods: Manufacturer of ArcCHECK phantom provides CT data set of phantom & recommends considering it as a homogeneous phantom with electron density (1.19 gm/cc or 282 HU) close to PMMA. We performed this study on Eclipse TPS (V13, VMS) & trueBEAM STx VMS Linac & ArcCHECK phantom (SNC). Plans were generated for 6MV photon beam, 20cm×20cm field size at isocentre & SPD (Source to phantom distance) of 86.7 cm to deliver 100cGy at isocentre. 3DVH software requires patients DICOM data generated by TPS & plan delivered on ArcCHECK phantom. Plans were generated in TPS by assigning different HU values to phantom. We analyzed gamma index & the dose profile for all plans along vertical down direction of beam’s central axis for Entry, Exit & Isocentre dose. Results: The global gamma passing rate (2% & 2mm) for manufacturer recommended HU value 282 was 96.3%. Detector entry, Isocentre & detector exit Doses were 1.9048 (1.9270), 1.00(1.0199) & 0.5078(0.527) Gy for TPS (Measured) respectively.The global gamma passing rate for electron density 1.1302 gm/cc was 98.6%. Detector entry, Isocentre & detector exit Doses were 1.8714 (1.8873), 1.00(0.9988) & 0.5211(0.516) Gy for TPS (Measured) respectively. Conclusion: Electron density value assigned by manufacturer does not hold true for every user. Proper modeling of electron density of ArcCHECK in TPS is essential to avoid systematic error in dose calculation of patient specific QA.

OSTI ID:
22648908
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-117: Analysis of the ArcCHECK Dosimetry Gamma Failure Using the 3DVH System
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22545243

Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA
Journal Article · Sat Jan 14 23:00:00 EST 2012 · Medical Physics · OSTI ID:22098738

Initial experience of ArcCHECK and 3DVH software for RapidArc treatment plan verification
Journal Article · Wed Oct 01 00:00:00 EDT 2014 · Medical Dosimetry · OSTI ID:22420859