Comparison of Dose Distributions With TG-43 and Collapsed Cone Convolution Algorithms Applied to Accelerated Partial Breast Irradiation Patient Plans
Journal Article
·
· International Journal of Radiation Oncology, Biology and Physics
- Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States)
- Department of Radiation Physics, The University of Texas Graduate School of Biomedical Sciences at Houston, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States)
Purpose: To compare the treatment plans for accelerated partial breast irradiation calculated by the new commercially available collapsed cone convolution (CCC) and current standard TG-43–based algorithms for 50 patients treated at our institution with either a Strut-Adjusted Volume Implant (SAVI) or Contura device. Methods and Materials: We recalculated target coverage, volume of highly dosed normal tissue, and dose to organs at risk (ribs, skin, and lung) with each algorithm. For 1 case an artificial air pocket was added to simulate 10% nonconformance. We performed a Wilcoxon signed rank test to determine the median differences in the clinical indices V90, V95, V100, V150, V200, and highest-dosed 0.1 cm{sup 3} and 1.0 cm{sup 3} of rib, skin, and lung between the two algorithms. Results: The CCC algorithm calculated lower values on average for all dose-volume histogram parameters. Across the entire patient cohort, the median difference in the clinical indices calculated by the 2 algorithms was <10% for dose to organs at risk, <5% for target volume coverage (V90, V95, and V100), and <4 cm{sup 3} for dose to normal breast tissue (V150 and V200). No discernable difference was seen in the nonconformance case. Conclusions: We found that on average over our patient population CCC calculated (<10%) lower doses than TG-43. These results should inform clinicians as they prepare for the transition to heterogeneous dose calculation algorithms and determine whether clinical tolerance limits warrant modification.
- OSTI ID:
- 22648772
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 5 Vol. 95; ISSN IOBPD3; ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-F-19A-01: APBI Brachytherapy Treatment Planning: The Impact of Heterogeneous Dose Calculations
SU-F-BRA-14: Optimization of Dosimetric Guidelines for Accelerated Partial Breast Irradiation (APBI) Using the Strut-Adjusted Volume Implant (SAVI)
Improvements in Critical Dosimetric Endpoints Using the Contura Multilumen Balloon Breast Brachytherapy Catheter to Deliver Accelerated Partial Breast Irradiation: Preliminary Dosimetric Findings of a Phase IV Trial
Journal Article
·
Sun Jun 15 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22402285
SU-F-BRA-14: Optimization of Dosimetric Guidelines for Accelerated Partial Breast Irradiation (APBI) Using the Strut-Adjusted Volume Implant (SAVI)
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22555220
Improvements in Critical Dosimetric Endpoints Using the Contura Multilumen Balloon Breast Brachytherapy Catheter to Deliver Accelerated Partial Breast Irradiation: Preliminary Dosimetric Findings of a Phase IV Trial
Journal Article
·
Fri Dec 31 23:00:00 EST 2010
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:21491532