Title: Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1]; ;  [1];  [2];  [1]
  1. Department of Radiation Oncology, Winship Cancer Institute of Emory University, Atlanta, Georgia (United States)
  2. Department of Biostatistics and Bioinformatics, Winship Cancer Institute of Emory University, Atlanta, Georgia (United States)

Purpose: To determine the dosimetric effects of rotational errors on target coverage using volumetric modulated arc therapy (VMAT) for multitarget stereotactic radiosurgery (SRS). Methods and Materials: This retrospective study included 50 SRS cases, each with 2 intracranial planning target volumes (PTVs). Both PTVs were planned for simultaneous treatment to 21 Gy using a single-isocenter, noncoplanar VMAT SRS technique. Rotational errors of 0.5°, 1.0°, and 2.0° were simulated about all axes. The dose to 95% of the PTV (D95) and the volume covered by 95% of the prescribed dose (V95) were evaluated using multivariate analysis to determine how PTV coverage was related to PTV volume, PTV separation, and rotational error. Results: At 0.5° rotational error, D95 values and V95 coverage rates were ≥95% in all cases. For rotational errors of 1.0°, 7% of targets had D95 and V95 values <95%. Coverage worsened substantially when the rotational error increased to 2.0°: D95 and V95 values were >95% for only 63% of the targets. Multivariate analysis showed that PTV volume and distance to isocenter were strong predictors of target coverage. Conclusions: The effects of rotational errors on target coverage were studied across a broad range of SRS cases. In general, the risk of compromised coverage increased with decreasing target volume, increasing rotational error and increasing distance between targets. Multivariate regression models from this study may be used to quantify the dosimetric effects of rotational errors on target coverage given patient-specific input parameters of PTV volume and distance to isocenter.

OSTI ID:
22644977
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 93, Issue 3; Other Information: Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-450: Dosimetric Impact of Rotational Error On Multiple-Target Intensity-Modulated Radiosurgery (IMRS) with Single-Isocenter
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22644977

SU-E-T-528: Robustness Evaluation for Fiducial-Based Accelerated Partial Breast Proton Therapy
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22644977

Volumetric Modulated Arc Therapy: Planning and Evaluation for Prostate Cancer Cases
Journal Article · Thu Apr 15 00:00:00 EDT 2010 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22644977