skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-F-T-179: Fast and Accurate Profile Acquisition for Proton Beam Using Multi-Ion Chamber Arrays

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4956316· OSTI ID:22642420
; ; ; ;  [1];  [2]
  1. Rutgers Cancer Institute of New Jersey, New Brunswick, NJ (United States)
  2. University of Wisconsin- Madison, Madison, WI (United States)

Purpose: Proton beam profile measurement is more time-consuming than photon beam. Due to the energy modulation during proton delivery, chambers have to move step-by-step instead of continuously. Multi-ion chamber arrays are appealing to this task since multiple measurements can be performed at once. However, their utilization suffers from sparse spatial resolution and potential intrinsic volume-averaging effect of the disk-shaped ion chambers. We proposed an approach to measure proton beam profiles accurately and efficiently. Methods: Mevion S250 proton system and IBA Matrixx ion chamber arrays were used in this study. Matrixx has interchamber distance of 7.62 mm, and chamber diameter of 4.5 mm. We measured the same beam profile by moving the Matrixx seven times with 1 mm each time along y axis. All 7 measurements were superimposed to get a “finer” profile with 1 mm spatial resolution. Coarser resolution profiles of 2 mm and 3 mm were also generated by using subsets of measurements. Those profiles were compared to the TPS calculated beam profile. Gamma analysis was performed for 2D dose maps to evaluate the difference to TPS dose plane. Results: Preliminary results showed a large discrepancy between the TPS calculated profile and the single measurement profile with 7.6 mm resolution. A good match could be achieved when the resolution reduced to 3 mm by adding one extra measurement. Gamma analysis for 2D dose map of a 10×10 field showed a passing rate (γ ≤ 1) of 90.6% using a 3% and 3mm criterion for single measurement, which increased to 92.3% for 2-measurement superimposition, and slightly further increased to 92.9% for 7-measurement superimposition. Conclusion: The results indicated that 2 measurements shifted by 3mm using Matrixx generated a smooth proton beam profile with good matching to Eclipse beam profile. We suggest using this 2-measurement approach in clinic for double scattering proton beam profile measurement.

OSTI ID:
22642420
Journal Information:
Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English