Title: SU-F-T-88: Field’s Metal for Electron Beam Inserts and Blocks

Abstract

Purpose: The aim of this work is to show that Field’s metal can be used as a viable alternative material to lead-based low-melting temperature alloys (e.g., Cerrobend) for electron beam inserts and blocks. The goal is to eliminate exposure risks associated with lead and cadmium. Methods: Field’s metal (51% Indium, 32.5% Bismuth, 16.5% Tin) is a low-melting point eutectic alloy, and is proposed as a lead-free replacement for Cadmium-free alloy (52.5% Bismuth, 32% Lead, 15.5% Tin). The experiments were done using a Varian 21EX linac equipped with 6, 9, 12, and 16 MeV beams. The 10×10 cone was used with 1.8cm thick insert made of Fields metal, Cd-Free alloy or open field. The transmitted radiation was measured using a diode at various depths in water. Block transmission factor (BTF) is determined as the ratio of maximum transmission radiation dose over the dose at dmax of the open field. Results: The data shows BTFs for both alloys are very similar, well below 5% for 6, 9, and 12 MeV beams. For 16 MeV, the BTF of the Fields metal is about 5.3%; the Cd-free metal 4.2%. The peak dose of the transmitted radiation occurs less than 1 cm below the watermore » surface, at a much shallower depth than the dmax of the electron beam in the open field. The beam profile shows the maximum block transmission is along the central axis of the beam. Conclusion: The present study shows that Fields metal can be used to make electron cutouts and blocks for 6, 9, 12, and 16 MeV beams. Because it has a similar density as other lead-based alloys, the same thickness of Fields metal would be adequate for shielding the electron beams. This means that it can be used as a replacement without modification to the existing block holder.« less

Authors:
; ; ;  [1]
  1. University Southern California, Los Angeles, CA (United States)
Publication Date:
OSTI Identifier:
22642336
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
43 PARTICLE ACCELERATORS; 60 APPLIED LIFE SCIENCES; BEAM PROFILES; CADMIUM ALLOYS; ELECTRON BEAMS; LEAD BASE ALLOYS; LINEAR ACCELERATORS; MEV RANGE 10-100; RADIATION DOSES; TRANSMISSION

Citation Formats

Ha, J, Gallogly, A, Cui, J, and Jennelle, R. SU-F-T-88: Field’s Metal for Electron Beam Inserts and Blocks. United States: N. p., 2016. Web. doi:10.1118/1.4956224.
Ha, J, Gallogly, A, Cui, J, & Jennelle, R. SU-F-T-88: Field’s Metal for Electron Beam Inserts and Blocks. United States. doi:10.1118/1.4956224.
Ha, J, Gallogly, A, Cui, J, and Jennelle, R. Wed . "SU-F-T-88: Field’s Metal for Electron Beam Inserts and Blocks". United States. doi:10.1118/1.4956224.
@article{osti_22642336,
title = {SU-F-T-88: Field’s Metal for Electron Beam Inserts and Blocks},
author = {Ha, J and Gallogly, A and Cui, J and Jennelle, R},
abstractNote = {Purpose: The aim of this work is to show that Field’s metal can be used as a viable alternative material to lead-based low-melting temperature alloys (e.g., Cerrobend) for electron beam inserts and blocks. The goal is to eliminate exposure risks associated with lead and cadmium. Methods: Field’s metal (51% Indium, 32.5% Bismuth, 16.5% Tin) is a low-melting point eutectic alloy, and is proposed as a lead-free replacement for Cadmium-free alloy (52.5% Bismuth, 32% Lead, 15.5% Tin). The experiments were done using a Varian 21EX linac equipped with 6, 9, 12, and 16 MeV beams. The 10×10 cone was used with 1.8cm thick insert made of Fields metal, Cd-Free alloy or open field. The transmitted radiation was measured using a diode at various depths in water. Block transmission factor (BTF) is determined as the ratio of maximum transmission radiation dose over the dose at dmax of the open field. Results: The data shows BTFs for both alloys are very similar, well below 5% for 6, 9, and 12 MeV beams. For 16 MeV, the BTF of the Fields metal is about 5.3%; the Cd-free metal 4.2%. The peak dose of the transmitted radiation occurs less than 1 cm below the water surface, at a much shallower depth than the dmax of the electron beam in the open field. The beam profile shows the maximum block transmission is along the central axis of the beam. Conclusion: The present study shows that Fields metal can be used to make electron cutouts and blocks for 6, 9, 12, and 16 MeV beams. Because it has a similar density as other lead-based alloys, the same thickness of Fields metal would be adequate for shielding the electron beams. This means that it can be used as a replacement without modification to the existing block holder.},
doi = {10.1118/1.4956224},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = {Wed Jun 15 00:00:00 EDT 2016},
month = {Wed Jun 15 00:00:00 EDT 2016}
}
  • Purpose: Rotational total skin electron irradiation (RTSEI) is used in the treatment of cutaneous T-cell lymphoma. Due to inter-film uniformity variations the dosimetry measurement of a large electron beam of a very low energy is challenging. This work provides a method to improve the accuracy of flatness and symmetry for a very large treatment field of low electron energy used in dual beam RTSEI. Methods: RTSEI is delivered by dual angles field a gantry of ±20 degrees of 270 to cover the upper and the lower halves of the patient body with acceptable beam uniformity. The field size is inmore » the order of 230cm in vertical height and 120 cm in horizontal width and beam energy is a degraded 6 MeV (6 mm of PMMA spoiler). We utilized parallel plate chambers, Gafchromic films and OSLDs as a measuring devices for absolute dose, B-Factor, stationary and rotational percent depth dose and beam uniformity. To reduce inter-film dosimetric variation we introduced a new specific correction method to analyze beam uniformity. This correction method uses some image processing techniques combining film value before and after radiation dose to compensate the inter-variation dose response differences among films. Results: Stationary and rotational depth of dose demonstrated that the Rp is 2 cm for rotational and the maximum dose is shifted toward the surface (3mm). The dosimetry for the phantom showed that dose uniformity reduced to 3.01% for the vertical flatness and 2.35% for horizontal flatness after correction thus achieving better flatness and uniformity. The absolute dose readings of calibrated films after our correction matched with the readings from OSLD. Conclusion: The proposed correction method for Gafchromic films will be a useful tool to correct inter-film dosimetric variation for the future clinical film dosimetry verification in very large fields, allowing the optimizations of other parameters.« less
  • Purpose: It is essential to assess the lung dose during TBI to reduce toxicity. Here we characterize the accuracy of the AAA and Acuros algorithms when using cerrobend lung shielding blocks at an extended distance for TBI. Methods: We positioned a 30×30×30 cm3 solid water slab phantom at 400 cm SSD and measured PDDs (Exradin A12 and PTW parallel plate ion chambers). A 2 cm thick, 10×10 cm2 cerrobend block was hung 2 cm in front of the phantom. This geometry was reproduced in the planning system for both AAA and Acuros. In AAA, the mass density of the cerrobendmore » block was forced to 9.38 g/cm3 and in Acuros it was forced to 8.0 g/cm3 (limited to selecting stainless steel). Three different relative electron densities (RED) were tested for each algorithm; 4.97, 6.97, and 8.97. Results: PDDs from both Acuros and AAA underestimated the delivered dose. AAA calculated that depth dose was higher for RED of 4.97 as compared to 6.97 and 8.97 but still lower than measured. There was no change in the percent depth dose with changing relative electron densities for Acuros. Conclusion: Care should be taken before using AAA or Acuros with cerrobend blocks as the planning system underestimates dose. Acuros limits the ability to modify RED when compared to AAA.« less
  • Purpose: Electron beam dosimetry requires high resolution data due to finite range that can be accomplished with small volume detectors. The small-field used in advance technologies in photon beam has created a market for microdetectors, however characteristics are significantly variable in photon beams and relatively unknown in electron beam that is investigated in this study. Methods: Among nearly 2 dozen microdetectors that have been investigated in small fields of photon beam, two popular detectors (microDiamond 60019 (PTW)) and W1 plastic scintillator detector (Standard Imaging)) that are tissue equivalent and have very small sensitive volume are selected. Electron beams from Varianmore » linear accelerators were used to investigate dose linearity dose rate dependence, energy dependence, depth dose and profiles in a reference condition in a water phantom. For W1 that has its own Supermax electrometer point by point measurements were performed. For microDiamond, a PTW-scanning tank was used for both scanning and point dose measurements. Results: W1 detector showed excellent dose linearity (r{sup 2} =1.0) from 5–500 MU either with variation of dose rate or beam energy. Similar findings were also observed for microdiamond with r{sup 2}=1.0. Percent variations in dose/MU for W1 and microDiamond were 0.2–1.1% and 0.4–1.2%, respectively among dose rate and beam energy. This variation was random for microDiamond, whereas it decreased with beam energy and dose rate for W1. The depth dose and profiles were within ±1 mm for both detectors. Both detectors did not show any energy dependence in electron beams. Conclusion: Both microDiamond and W1 detectors provided superior characteristics of beam parameters in electron beam including dose, dose rate linearity and energy independence. Both can be used in electron beam except W1 require point by point measurements and microdiamond requires 1500 MU for initial quenching.« less
  • Purpose: To estimate the dosimetric potential of mixed beam radiotherapy (MBRT) by using a single process optimizing the shape and weight of photon and electron apertures simultaneously based on Monte Carlo beamlet dose distributions. Methods: A simulated annealing based direct aperture optimization capable to perform simultaneous optimization was developed to generate treatment plans for MERT, photon-IMRT and MBRT. Both photon and electron apertures are collimated with the photon-MLC and are delivered in a segmented manner. For dosimetric comparison and for investigating the dependency on the number of apertures, photon-IMRT, MERT and MBRT plans were generated for an academic case consistingmore » of a water phantom containing two shallow PTVs differing in the maximal depth of 5 and 7 cm, respectively and two OARs in distal and lateral direction to the PTVs. Results: For the superficial PTV, the dose homogeneity (V95%–V107%) and the mean dose (in percent of the prescribed dose) to the distal and the lateral OARs of the MBRT plan (94.9%, 16.9%, 17.8%) are superior or comparable to those for the MERT (74%, 18.4%, 15.4%) and the photon-IMRT plan (89.4%, 20.8%, 24.7%). For the enlarged PTV, the dosimetric superiority of MBRT compared to MERT and photon-IMRT is even more pronounced. Furthermore, an MBRT plan with 12 electron and 10 photon apertures lead to an objective function value 38% lower than that of a photon-IMRT plan with 40 apertures. Conclusion: The results of simultaneous optimization for MBRT are promising with regards to further OAR sparing and improved dose coverage to the PTV compared to photon-IMRT and MERT. Especially superficial targets with deeper subparts (>5 cm) could substantially benefit. Moreover, MBRT seems to be a possible solution of two downsides of photon-IMRT, namely the extended low dose bath and the requirement of numerous apertures. This work was supported by Varian Medical Systems. This work was supported by Varian Medical Systems.« less
  • Purpose: Validation of iBEAM™ evo couch-top for different relative electron density (RED) combination during photon beam dose calculation in Monaco− TPS. Methods: The iBEAM™ evo couch-top has two layers:outer carbon fiber (CF) and inner foam core (FC). To study the beam intensity attenuation of couch-top, measured doses were compared with doses calculated for different REDs. Measurements were performed in solid water phantom with PTW-0.125cc ion-chamber positioned at center of the phantom with 5.3cm thickness slabs placed above and below the chamber. Similarly, in TPS, iBEAM™ evo couch-top was simulated and doses were calculated for different RED combinations (0.2CF-0.2FC, 0.4CF-0.2FC, 0.6CF-0.2FC,more » 0.8CF-0.2FC, and 1.0CF-0.2FC) by using Monte Carlo dose calculation algorithm in Monaco TPS (V5.1). Doses were measured for every 10 degree gantry angle separation, 10×10cm{sup 2} field size and 6MV photons. Then, attenuation is defined as the ratio of output at posterior gantry angle to output of its opposed anterior gantry angle (e.g.225°/45°). output fluctuation with different gantry angle was within ±0.21%. To confirm above results, dose-planes were measured for five pelvic VMAT plans (360°arc) in PTW two-dimensional array and compared with different calculated dose-planes of above-mentioned couch REDs. Gamma pass rates<1.00) were analyzed for 3%/2mm criteria. Results: Measured and calculated attenuation was in good agreement for the RED combination of 0.2CF-0.2FC and difference was within ±0.515%. However, other density combination showed difference of ±0.9841%, ±1.667%, ±2.9241% and ±2.8832% for 0.4CF-0.2FC, 0.6CF-0.2FC, 0.8CF-0.2FC, and 1.0CF-0.2FC, respectively. Maximum couch-top attenuation was observed at 110°–120° and 240°–250° and decreases linearly as the gantry angle approaches 180°. Moreover, gamma pass rate confirmed the above results and showed maximum pass rate of 96.23% for 0.2CF-0.2FC, whereas others were 95.72%, 95.12%, 94.31% and 93.24%. Conclusion: RED value of 0.2CF-0.2FC was found to be suitable for accurate couch-top modeling for 6MV photon beam Monte Carlo calculations in Monaco TPS.« less