Title: SU-F-J-73: Simple Approach for Quantification of Metal Artifact Reduction Capabalities of Dual-Energy CT

Abstract

Purpose: To present a simple method for quantification of dual-energy CT metal artifact reduction capabilities Methods: A phantom was constructed from solid water and a steel cylinder. Solid water is commonly used for radiotherapy QA, while steel cylinders are readily available in hardware stores. The phantom was scanned on Siemens Somatom 64-slice dual-energy CT system. Three CTs were acquired at energies of 80kV (low), 120kV (nominal), and 140kV (high). The low and high energy acquisitions were used to generate dual-energy (DE) monoenergetic image sets, which also utilized metal artifact reduction algorithm (Maris). Several monoenergetic DE image sets, ranging from 70keV to 190keV were generated. The size of the metal artifact was measured by two different approaches. The first approach measured the distance from the center of the steel cylinder to a location with nominal (undisturbed by metal) HU value for the 120kV, DE 70keV, and DE 190keV image sets. In the second approach, the distance from the center of the cylinder to the edge of the air pocket for the above mentioned three image sets was measured. Results: The DE 190keV synthetic image set demonstrated the largest reduction of the metal artifacts. The size of the artifact was more thanmore » three times the actual size of the milled hole in the solid water in the DE 190keV, as compared to more than 7.5 times larger as estimated from the 120kV uncorrected image Conclusion: A simple phantom for quantification of dual-energy CT metal artifact reduction capabilities was presented. This inexpensive phantom can be easily built from components available in every radiation oncology department. It allows quick assessment and quantification of the properties of different metal artifact reduction algorithms, available on modern dual-energy CT scanners.« less

Authors:
; ; ;  [1]
  1. University of Miami, Miami, FL (United States)
Publication Date:
OSTI Identifier:
22632202
Resource Type:
Journal Article
Resource Relation:
Journal Name: Medical Physics; Journal Volume: 43; Journal Issue: 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA)
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; 61 RADIATION PROTECTION AND DOSIMETRY; ALGORITHMS; COMPUTERIZED TOMOGRAPHY; DISTANCE; IMAGES; PHANTOMS; RADIOTHERAPY; STEELS

Citation Formats

Lamichhane, N, Padgett, K, Li, X, and Mihaylov, I. SU-F-J-73: Simple Approach for Quantification of Metal Artifact Reduction Capabalities of Dual-Energy CT. United States: N. p., 2016. Web. doi:10.1118/1.4955981.
Lamichhane, N, Padgett, K, Li, X, & Mihaylov, I. SU-F-J-73: Simple Approach for Quantification of Metal Artifact Reduction Capabalities of Dual-Energy CT. United States. doi:10.1118/1.4955981.
Lamichhane, N, Padgett, K, Li, X, and Mihaylov, I. 2016. "SU-F-J-73: Simple Approach for Quantification of Metal Artifact Reduction Capabalities of Dual-Energy CT". United States. doi:10.1118/1.4955981.
@article{osti_22632202,
title = {SU-F-J-73: Simple Approach for Quantification of Metal Artifact Reduction Capabalities of Dual-Energy CT},
author = {Lamichhane, N and Padgett, K and Li, X and Mihaylov, I},
abstractNote = {Purpose: To present a simple method for quantification of dual-energy CT metal artifact reduction capabilities Methods: A phantom was constructed from solid water and a steel cylinder. Solid water is commonly used for radiotherapy QA, while steel cylinders are readily available in hardware stores. The phantom was scanned on Siemens Somatom 64-slice dual-energy CT system. Three CTs were acquired at energies of 80kV (low), 120kV (nominal), and 140kV (high). The low and high energy acquisitions were used to generate dual-energy (DE) monoenergetic image sets, which also utilized metal artifact reduction algorithm (Maris). Several monoenergetic DE image sets, ranging from 70keV to 190keV were generated. The size of the metal artifact was measured by two different approaches. The first approach measured the distance from the center of the steel cylinder to a location with nominal (undisturbed by metal) HU value for the 120kV, DE 70keV, and DE 190keV image sets. In the second approach, the distance from the center of the cylinder to the edge of the air pocket for the above mentioned three image sets was measured. Results: The DE 190keV synthetic image set demonstrated the largest reduction of the metal artifacts. The size of the artifact was more than three times the actual size of the milled hole in the solid water in the DE 190keV, as compared to more than 7.5 times larger as estimated from the 120kV uncorrected image Conclusion: A simple phantom for quantification of dual-energy CT metal artifact reduction capabilities was presented. This inexpensive phantom can be easily built from components available in every radiation oncology department. It allows quick assessment and quantification of the properties of different metal artifact reduction algorithms, available on modern dual-energy CT scanners.},
doi = {10.1118/1.4955981},
journal = {Medical Physics},
number = 6,
volume = 43,
place = {United States},
year = 2016,
month = 6
}
  • Purpose: The purpose of this abstract was to evaluate the performance of commercial strategies for artifact reduction in radiation oncology settings. The iterative metal artifact reduction (Siemens iMAR) algorithm and monoenergetic virtual datasets reconstructed from dual energy scans are compared side-by-side in their ability to image in the presence of metal inserts. Methods: A CIRS ATOM Dosimetry Verification Phantom was scanned with and without a metal insert on a SOMATOM Definition AS dual energy scanner. Images with the metal insert were reconstructed with (a) a tradition single energy CT scan with the iMAR option implemented, using different artifact reduction settingsmore » and (b) a monoenergetic scan calculated from dual energy scans by recovering differences in the energy-dependence of the attenuation coefficients of different materials and then creating a virtual monoenergetic scan from these coefficients. The iMAR and monoenergetic scans were then compared with the metal-free scan to assess changes in HU numbers and noise within a region around the metal insert. Results: Both the iMAR and dual energy scans reduced artifacts produced by the metal insert. However the iMAR results are dependent of the selected algorithm settings, with a mean HU difference ranging from 0.65 to 90.40 for different options. The mean differences without the iMAR correction were 38.74. When using the dual energy scan, the mean differences were 4.53, that is however attributed to increased noise and not artifacts, as the dual energy scan had the lowest skewness (2.52) compared to the iMAR scans (ranging from 3.90 to 4.88) and the lowest kurtosis (5.72 for dual energy, range of 18.19 to 27.36 for iMAR). Conclusion: Both approaches accurately recovered HU numbers, however the dual energy method provided smaller residual artifacts.« less
  • Purpose: In this article, an approach to metal artifact reduction is proposed that is practical for clinical use in radiation therapy. It is based on a new interpolation scheme of the projections associated with metal implants in helical computed tomography (CT) scanners. Methods and Materials: A three-step approach was developed consisting of an automatic algorithm for metal implant detection, a correction algorithm for helical projections, and a new, efficient algorithm for projection interpolation. The modified raw projection data are transferred back to the CT scanner device where CT slices are regenerated using the built-in reconstruction operator. The algorithm was testedmore » on a CT calibration phantom in which the density of inserted objects are known and on clinical prostate cases with two hip prostheses. The results are evaluated using the CT number and shape of the objects. Results: The validations on a CT calibration phantom with various inserts of known densities show that the algorithm improved the overall image quality by restoring the shape and the representative CT number of the objects in the image. For the clinical hip replacement cases, a large fraction of the bladder, rectum, and prostate that were not visible on the original CT slices were recovered using the algorithm. Precise contouring of the target volume was thus feasible. Without this enhancement, physicians would have drawn bigger margins to be sure to include the target and, at the same time, could have prescribed a lower dose to keep the same level of normal tissue toxicity. Conclusions: In both phantom experiment and patient studies, the algorithm resulted in significant artifact reduction with increases in the reliability of planning procedure for the case of metallic hip prostheses. This algorithm is now clinically used as a preprocessing before treatment planning for metal artifact reduction.« less
  • Purpose: In computed tomography imaging metal objects in the region of interest introduce inconsistencies during data acquisition. Reconstructing these data leads to an image in spatial domain including star-shaped or stripe-like artifacts. In order to enhance the quality of the resulting image the influence of the metal objects can be reduced. Here, a metal artifact reduction (MAR) approach is proposed that is based on a recomputation of the inconsistent projection data using a fully three-dimensional Fourier-based interpolation. The success of the projection space restoration depends sensitively on a sensible continuation of neighboring structures into the recomputed area. Fortunately, structural informationmore » of the entire data is inherently included in the Fourier space of the data. This can be used for a reasonable recomputation of the inconsistent projection data. Methods: The key step of the proposed MAR strategy is the recomputation of the inconsistent projection data based on an interpolation using nonequispaced fast Fourier transforms (NFFT). The NFFT interpolation can be applied in arbitrary dimension. The approach overcomes the problem of adequate neighborhood definitions on irregular grids, since this is inherently given through the usage of higher dimensional Fourier transforms. Here, applications up to the third interpolation dimension are presented and validated. Furthermore, prior knowledge may be included by an appropriate damping of the transform during the interpolation step. This MAR method is applicable on each angular view of a detector row, on two-dimensional projection data as well as on three-dimensional projection data, e.g., a set of sequential acquisitions at different spatial positions, projection data of a spiral acquisition, or cone-beam projection data. Results: Results of the novel MAR scheme based on one-, two-, and three-dimensional NFFT interpolations are presented. All results are compared in projection data space and spatial domain with the well-known one-dimensional linear interpolation strategy. Conclusions: In conclusion, it is recommended to include as much spatial information into the recomputation step as possible. This is realized by increasing the dimension of the NFFT. The resulting image quality can be enhanced considerably.« less
  • Purpose: To investigate two strategies for reducing dose calculation errors near metal implants: use of CT metal artifact reduction methods and implementation of metal-based energy deposition kernels in the convolution/superposition (C/S) method. Methods: Radiochromic film was used to measure the dose upstream and downstream of titanium and Cerrobend implants. To assess the dosimetric impact of metal artifact reduction methods, dose calculations were performed using baseline, uncorrected images and metal artifact reduction Methods: Philips O-MAR, GE’s monochromatic gemstone spectral imaging (GSI) using dual-energy CT, and GSI imaging with metal artifact reduction software applied (MARs).To assess the impact of metal kernels, titaniummore » and silver kernels were implemented into a commercial collapsed cone C/S algorithm. Results: The CT artifact reduction methods were more successful for titanium than Cerrobend. Interestingly, for beams traversing the metal implant, we found that errors in the dimensions of the metal in the CT images were more important for dose calculation accuracy than reduction of imaging artifacts. The MARs algorithm caused a distortion in the shape of the titanium implant that substantially worsened the calculation accuracy. In comparison to water kernel dose calculations, metal kernels resulted in better modeling of the increased backscatter dose at the upstream interface but decreased accuracy directly downstream of the metal. We also found that the success of metal kernels was dependent on dose grid size, with smaller calculation voxels giving better accuracy. Conclusion: Our study yielded mixed results, with neither the metal artifact reduction methods nor the metal kernels being globally effective at improving dose calculation accuracy. However, some successes were observed. The MARs algorithm decreased errors downstream of Cerrobend by a factor of two, and metal kernels resulted in more accurate backscatter dose upstream of metals. Thus, these two strategies do have the potential to improve accuracy for patients with metal implants in certain scenarios. This work was supported by Public Health Service grants CA 180803 and CA 10953 awarded by the National Cancer Institute, United States of Health and Human Services, and in part by Mobius Medical Systems.« less
  • Purpose: The empirical dual energy calibration (EDEC) method corrects for beam-hardening artifacts, but shows limited performance on metal artifact correction. In this work, we propose an adaptive dual energy calibration (ADEC) method to correct for metal artifacts. Methods: The empirical dual energy calibration (EDEC) method corrects for beam-hardening artifacts, but shows limited performance on metal artifact correction. In this work, we propose an adaptive dual energy calibration (ADEC) method to correct for metal artifacts. Results: Highly attenuating copper rods cause severe streaking artifacts on standard CT images. EDEC improves the image quality, but cannot eliminate the streaking artifacts. Compared tomore » EDEC, the proposed ADEC method further reduces the streaking resulting from metallic inserts and beam-hardening effects and obtains material decomposition images with significantly improved accuracy. Conclusion: We propose an adaptive dual energy calibration method to correct for metal artifacts. ADEC is evaluated with the Shepp-Logan phantom, and shows superior metal artifact correction performance. In the future, we will further evaluate the performance of the proposed method with phantom and patient data.« less