skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Partially coherent wavefront propagation simulations: Mirror and monochromator crystal quality assessment

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4952885· OSTI ID:22608398
;  [1];  [2]
  1. NSLS-II Photon Science Division, Brookhaven National Laboratory, Upton NY (United States)
  2. RadiaSoft LLC, 1348 Redwood Ave., Boulder, CO 80304-1146 (United States)

We have applied fully-and partially-coherent synchrotron radiation wavefront propagation simulations, implemented in the “Synchrotron Radiation Workshop” (SRW) computer code, to analyse the effects of imperfect mirrors and monochromator at the Coherent Hard X-ray beamline. This beamline is designed for X-ray Photon Correlation Spectroscopy, a technique that heavily relies on the partial coherence of the X-ray beam and benefits from a careful preservation of the X-ray wavefront. We present simulations and a comparison with the measured beam profile at the sample position, which show the impact of imperfect optics on the wavefront.

OSTI ID:
22608398
Journal Information:
AIP Conference Proceedings, Vol. 1741, Issue 1; Conference: SRI2015: 12. international conference on synchrotron radiation instrumentation, New York, NY (United States), 6-10 Jul 2015; Other Information: (c) 2016 Author(s); Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English