Verification of high efficient broad beam cold cathode ion source
- Physics Department, Faculty of Science, Helwan University, Cairo (Egypt)
- Radiation Physics Department, National Center for Radiation Research and Technology (NCRRT), Atomic Energy Authority (AEA), Cairo (Egypt)
An improved form of cold cathode ion source has been designed and constructed. It consists of stainless steel hollow cylinder anode and stainless steel cathode disc, which are separated by a Teflon flange. The electrical discharge and output characteristics have been measured at different pressures using argon, nitrogen, and oxygen gases. The ion exit aperture shape and optimum distance between ion collector plate and cathode disc are studied. The stable discharge current and maximum output ion beam current have been obtained using grid exit aperture. It was found that the optimum distance between ion collector plate and ion exit aperture is equal to 6.25 cm. The cold cathode ion source is used to deposit aluminum coating layer on AZ31 magnesium alloy using argon ion beam current which equals 600 μA. Scanning electron microscope and X-ray diffraction techniques used for characterizing samples before and after aluminum deposition.
- OSTI ID:
- 22597713
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 8 Vol. 87; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Electron emission from glow-discharge cathode materials due to neon and argon ion bombardment
Optimization of Uranium Molecular Deposition for Alpha-Counting Sources
Related Subjects
75 CONDENSED MATTER PHYSICS
SUPERCONDUCTIVITY AND SUPERFLUIDITY
ALUMINIUM
ANODES
APERTURES
ARGON
ARGON IONS
BEAM CURRENTS
CATHODES
ELECTRIC DISCHARGES
ION BEAMS
ION SOURCES
MAGNESIUM ALLOYS
NITROGEN
OXYGEN
SCANNING ELECTRON MICROSCOPY
STAINLESS STEELS
TEFLON
X-RAY DIFFRACTION