Accuracy of UTE-MRI-based patient setup for brain cancer radiation therapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4938266· OSTI ID:22579811
; ; ; ; ; ; ; ; ;  [1]; ;  [2]
  1. Department of Radiation Oncology, University of California, Los Angeles, California 90095 (United States)
  2. Department of Radiological Sciences, University of California, Los Angeles, California 90095 (United States)

Purpose: Radiation therapy simulations solely based on MRI have advantages compared to CT-based approaches. One feature readily available from computed tomography (CT) that would need to be reproduced with MR is the ability to compute digitally reconstructed radiographs (DRRs) for comparison against on-board radiographs commonly used for patient positioning. In this study, the authors generate MR-based bone images using a single ultrashort echo time (UTE) pulse sequence and quantify their 3D and 2D image registration accuracy to CT and radiographic images for treatments in the cranium. Methods: Seven brain cancer patients were scanned at 1.5 T using a radial UTE sequence. The sequence acquired two images at two different echo times. The two images were processed using an in-house software to generate the UTE bone images. The resultant bone images were rigidly registered to simulation CT data and the registration error was determined using manually annotated landmarks as references. DRRs were created based on UTE-MRI and registered to simulated on-board images (OBIs) and actual clinical 2D oblique images from ExacTrac™. Results: UTE-MRI resulted in well visualized cranial, facial, and vertebral bones that quantitatively matched the bones in the CT images with geometric measurement errors of less than 1 mm. The registration error between DRRs generated from 3D UTE-MRI and the simulated 2D OBIs or the clinical oblique x-ray images was also less than 1 mm for all patients. Conclusions: UTE-MRI-based DRRs appear to be promising for daily patient setup of brain cancer radiotherapy with kV on-board imaging.

OSTI ID:
22579811
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 1 Vol. 43; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-E-J-90: MRI-Based Treatment Simulation and Patient Setup for Radiation Therapy of Brain Cancer
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22325245

SU-G-JeP2-08: Image-Guided Radiation Therapy Using Synthetic CTs in Brain Cancer
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649374

Magnetic Resonance–Based Automatic Air Segmentation for Generation of Synthetic Computed Tomography Scans in the Head Region
Journal Article · Sun Nov 01 00:00:00 EDT 2015 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22644972