Accuracy evaluation of the optical surface monitoring system on EDGE linear accelerator in a phantom study
Frameless stereotactic radiosurgery (SRS) requires dedicated systems to monitor the patient position during the treatment to avoid target underdosage due to involuntary shift. The optical surface monitoring system (OSMS) is here evaluated in a phantom-based study. The new EDGE linear accelerator from Varian (Varian, Palo Alto, CA) integrates, for cranial lesions, the common cone beam computed tomography (CBCT) and kV-MV portal images to the optical surface monitoring system (OSMS), a device able to detect real-time patient's face movements in all 6 couch axes (vertical, longitudinal, lateral, rotation along the vertical axis, pitch, and roll). We have evaluated the OSMS imaging capability in checking the phantoms' position and monitoring its motion. With this aim, a home-made cranial phantom was developed to evaluate the OSMS accuracy in 4 different experiments: (1) comparison with CBCT in isocenter location, (2) capability to recognize predefined shifts up to 2° or 3 cm, (3) evaluation at different couch angles, (4) ability to properly reconstruct the surface when the linac gantry visually block one of the cameras. The OSMS system showed, with a phantom, to be accurate for positioning in respect to the CBCT imaging system with differences of 0.6 ± 0.3 mm for linear vector displacement, with a maximum rotational inaccuracy of 0.3°. OSMS presented an accuracy of 0.3 mm for displacement up to 1 cm and 1°, and 0.5 mm for larger displacements. Different couch angles (45° and 90°) induced a mean vector uncertainty < 0.4 mm. Coverage of 1 camera produced an uncertainty < 0.5 mm. Translations and rotations of a phantom can be accurately detect with the optical surface detector system.
- OSTI ID:
- 22577880
- Journal Information:
- Medical Dosimetry, Vol. 41, Issue 2; Other Information: Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-G-JeP4-04: Commissioning and Acceptance Testing of Optical Surface Monitoring System On TrueBEAM STx as Per Task Group 147
SU-E-J-13: Six Degree of Freedom Image Fusion Accuracy for Cranial Target Localization On the Varian Edge Stereotactic Radiosurgery System: Comparison Between 2D/3D and KV CBCT Image Registration