Dosimetric feasibility of an “off-target isocenter” technique for cranial intensity-modulated radiosurgery
Journal Article
·
· Medical Dosimetry
To evaluate the dosimetric effect of placing the isocenter away from the planning target volume (PTV) on intensity-modulated radiosurgery (IMRS) plans to treat brain lesions. A total of 15 patients who received cranial IMRS at our institution were randomly selected. Each patient was treated with an IMRS plan designed with the isocenter located at the target center (plan A). A second off-target isocenter plan (plan B) was generated for each case. In all the plans,100% of the prescription dose covered 99% of the target volume. The plans A and B were compared for the target dosage (conformity index [CI] and homogeneity index) and organs-at-risk (OAR) dose sparing. Peripheral dose falloff was compared by using the metrics volume of normal brain receiving more than 12-Gy dose (V12) and CI at the level of the 50% of the prescription dose (CI 50%). The values found for each metric (plan B vs plan A) were (mean ± standard deviation [SD]) as follows—CI: 1.28 ± 0.15 vs 1.28 ± 0.15, p = 0.978; homogeneity index (HI): 1.29 ± 0.14 vs 1.34 ± 0.17, p = 0.079; maximum dose to the brainstem: 2.95 ± 2.11 vs 2.89 ± 1.88 Gy, p = 0.813; maximum dose to the optical pathway: 2.65 ± 4.18 vs 2.44 ± 4.03 Gy, p = 0.195; and maximum dose to the eye lens: 0.33 ± 0.73 vs 0.33 ± 0.53 Gy, p = 0.970. The values of the peripheral dose falloff were (plan B vs plan A) as follows—V12: 5.98 ± 4.95 vs 6.06 ± 4.92 cm{sup 3}, p = 0.622, and CI 50%: 6.08 ± 2.77 vs 6.28 ± 3.01, p = 0.119. The off-target isocenter solution resulted in dosimetrically comparable plans as the center-target isocenter technique, by avoiding the risk of gantry-couch collision during the cone beam computed tomography (CBCT) acquisition.
- OSTI ID:
- 22577837
- Journal Information:
- Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 4 Vol. 40; ISSN 0958-3947
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-F-T-510: Single Isocenter Radiosurgery for the Simultaneous Treatment of Multiple Brain Metastases: Volumetric Modulated Arc Therapy Or Dynamic Conformal Arc?
The Importance of the Conformality, Heterogeneity, and Gradient Indices in Evaluating Gamma Knife Radiosurgery Treatment Plans for Intracranial Meningiomas
Dosimetric and Clinical Analysis of Spatial Distribution of the Radiation Dose in Gamma Knife Radiosurgery for Vestibular Schwannoma
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22649097
The Importance of the Conformality, Heterogeneity, and Gradient Indices in Evaluating Gamma Knife Radiosurgery Treatment Plans for Intracranial Meningiomas
Journal Article
·
Wed Aug 01 00:00:00 EDT 2012
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:22149382
Dosimetric and Clinical Analysis of Spatial Distribution of the Radiation Dose in Gamma Knife Radiosurgery for Vestibular Schwannoma
Journal Article
·
Mon Nov 14 23:00:00 EST 2011
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:22054420