skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Modification of a Superficial X-Ray Therapy Machine for Rectal Contact Therapy

Journal Article · · Medical Dosimetry

X-ray therapy of superficial rectal cancers using a hand-held 50 kV contact unit (Philips RT-50) in a technique first described by Papillon had reached a point of widening clinical acceptability when the manufacturer of this equipment discontinued its production. To pursue this endocavitary approach to rectal therapy, technical modifications have to be made to conventional superficial x-ray therapy machines. Advantages over the original Papillon method include remote viewing of the lesion through the proctoscopic cone and a lower radiation exposure for the operator. We have evaluated a Bucky Combination Therapy Unit under conditions in which the operating voltage (65 kV), target skin distance (23.6 cm), and added filtration (0.39 mm Al) were selected in order to match as closely as possible the beam penetration characteristics of the “standard” (Papillon) technique. With this equipment, the thermal characteristics of the tube anode and housing limit the amount of radiation that can be delivered before a “rest period” for the machine is needed. In practice, 3 minutes of irradiation at an exposure rate of 500 R/min can be performed followed by an interval of 3 minutes before irradiation can be resumed.

OSTI ID:
22577828
Journal Information:
Medical Dosimetry, Vol. 19, Issue 1; Other Information: Copyright (c) 1994 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
Country of Publication:
United States
Language:
English

Similar Records

Radiation protection considerations for endocavitary X ray units
Journal Article · Sun Dec 01 00:00:00 EST 1985 · Int. J. Radiat. Oncol., Biol. Phys.; (United States) · OSTI ID:22577828

Design and dosimetric characteristics of a new endocavitary contact radiotherapy system using an electronic brachytherapy source
Journal Article · Thu Nov 15 00:00:00 EST 2012 · Medical Physics · OSTI ID:22577828

Endocavitary irradiation for rectal cancer and villous adenomas
Journal Article · Mon Feb 01 00:00:00 EST 1988 · Int. J. Radiat. Oncol., Biol. Phys.; (United States) · OSTI ID:22577828