TU-B-304-00: The Aftermath of TG-142
Although published in 2009, the AAPM TG-142 report on accelerator quality assurance still proves a challenge for full clinical implementation. The choice of methodologies to satisfy TG-142 requirements is critical to a successful application. Understanding the philosophy of TG-142 can help in creating an institution-specific QA practice that is both efficient and effective. The concept of maintaining commissioned beam profiles is still found confusing. The physicist must also consider technologies not covered by TG-142 (i.e. arc therapy techniques). On the horizon is TG-198 report on implementing TG-142. Although the community still lacks a final TG-100 report, performing a failure-mode -and-effects analysis and statistical process control analysis to determine the institution-specific clinical impact of each TG-142 test may be useful for identifying trends for pro-active surveillance. Learning Objectives: To better understand the confusing and controversial aspects of TG-142. To understand what is still missing from TG-142 and how to account for these tests in clinical practice To describe which QA tests in TG-142 yield the largest potential clinical result if not discovered.
- OSTI ID:
- 22563028
- Journal Information:
- Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
TU-B-304-02: Quantitative FMEA of TG-142
WE-PIS-Exhibit Hall-01: Tools for TG-142 Linac Imaging QA II