SU-E-T-800: Verification of Acurose XB Dose Calculation Algorithm at Air Cavity-Tissue Interface Using Film Measurement for Small Fields of 6-MV Flattening Filter-Free Beams
- The Catholic University of Korea, Seoul (Korea, Republic of)
- Seoul National University Bundang Hospital, Seongnam, Kyeonggi-do (Korea, Republic of)
Purpose: To verify the dose accuracy of Acuros XB (AXB) dose calculation algorithm at air-tissue interface using inhomogeneous phantom for 6-MV flattening filter-free (FFF) beams. Methods: An inhomogeneous phantom included air cavity was manufactured for verifying dose accuracy at the air-tissue interface. The phantom was composed with 1 and 3 cm thickness of air cavity. To evaluate the central axis doses (CAD) and dose profiles of the interface, the dose calculations were performed for 3 × 3 and 4 × 4 cm{sup 2} fields of 6 MV FFF beams with AAA and AXB in Eclipse treatment plainning system. Measurements in this region were performed with Gafchromic film. The root mean square errors (RMSE) were analyzed with calculated and measured dose profile. Dose profiles were divided into inner-dose profile (>80%) and penumbra (20% to 80%) region for evaluating RMSE. To quantify the distribution difference, gamma evaluation was used and determined the agreement with 3%/3mm criteria. Results: The percentage differences (%Diffs) between measured and calculated CAD in the interface, AXB shows more agreement than AAA. The %Diffs were increased with increasing the thickness of air cavity size and it is similar for both algorithms. In RMSEs of inner-profile, AXB was more accurate than AAA. The difference was up to 6 times due to overestimation by AAA. RMSEs of penumbra appeared to high difference for increasing the measurement depth. Gamma agreement also presented that the passing rates decreased in penumbra. Conclusion: This study demonstrated that the dose calculation with AXB shows more accurate than with AAA for the air-tissue interface. The 2D dose distributions with AXB for both inner-profile and penumbra showed better agreement than with AAA relative to variation of the measurement depths and air cavity sizes.
- OSTI ID:
- 22555165
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-E-T-339: Dosimetric Verification of Acuros XB Dose Calculation Algorithm On An Air Cavity for 6-MV Flattening Filter-Free Beam
SU-F-SPS-01: Accuracy of the Small Field Dosimetry Using Acuros XB and AAA Dose Calculation Algorithms of Eclipse Treatment Planning System Within and Beyond Heterogeneous Media for Trubeam 2.0 Unit
Dosimetric impact of Acuros XB deterministic radiation transport algorithm for heterogeneous dose calculation in lung cancer
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22548393
SU-F-SPS-01: Accuracy of the Small Field Dosimetry Using Acuros XB and AAA Dose Calculation Algorithms of Eclipse Treatment Planning System Within and Beyond Heterogeneous Media for Trubeam 2.0 Unit
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22624418
Dosimetric impact of Acuros XB deterministic radiation transport algorithm for heterogeneous dose calculation in lung cancer
Journal Article
·
Wed May 15 00:00:00 EDT 2013
· Medical Physics
·
OSTI ID:22130613