Title: SU-E-T-311: Dosimetric Comparison of Volumetric Modulated Arc Therapy Plans for Preoperative Radiotherapy Rectal Cancer Using Flattening Filter-Free and Flattening Filter Modes

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924672· OSTI ID:22548367
; ; ;  [1]
  1. Cancer Hospital of Shantou University Medical College, Shantou, Guangdong (China)

Purpose: To compare the dosimetric difference of volumetric modulated arc therapy(VMAT) for preoperative radiotherapy rectal cancer using 6MV X-ray flattening filter free(FFF) and flattening filter(FF) modes. Methods: FF-VMAT and FFF-VMAT plans were designed to 15 rectal cancer patients with preoperative radiotherapy by planning treatment system(Eclipse 10.0),respectively. Dose prescription was 50 Gy in 25 fractions. All plans were normalized to 50 Gy to 95% of PTV. The Dose Volume Histogram (DVH), target and risk organ doses, conformity indexes (CI), homogeneity indexes (HI), low dose volume of normal tissue(BP), monitor units(MU) and treatment time (TT) were compared between the two kinds of plans. Results: FF-VMAT provided the lower Dmean, V105, HI, and higher CI as compared with FFF-VMAT. The small intestine of D5, Bladder of D5, Dmean, V40, V50, L-femoral head of V40, R-femoral head of Dmean were lower in FF-VMAT than in FFF-VMAT. FF-VMAT had higher BP of V5, but no significantly different of V10, V15, V20, V30 as compared with FFF-VMAT. FF-VMAT reduceed the monitor units(MU) by 21%(P<0.05), as well as the treatment time(TT) was no significantly different(P>0.05), as compared with FFF-VMAT. Conclusion: The plan qualities of FF and FFF VMAT plans were comparable and both clinically acceptable. FF-VMAT as compared with FFF-VMAT, showing better target coverage, some of OARs sparing, the MUs of FFF-VMAT were higher than FF-VMAT, yet were delivered within the same time. This work was supported by the Medical Scientific Research Foundation of Guangdong Procvince (A2014455 to Changchun Ma)

OSTI ID:
22548367
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English