SU-E-T-100: An Approach to Improving the Dynamic Delivery Accuracy for Breast IMRT

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924461· OSTI ID:22545228
 [1]; ;  [1]; ;  [2]
  1. Grand River Hospital, Kitchener, ON (Canada)
  2. University of Waterloo, Waterloo, ON (Canada)

Purpose: A method is established to improve the accuracy of the IMRT dose delivery for the treatment of breast and chest wall tumors when the higher dose gradient is delivered at the end of the leaf sequence. Methods: Dynamic MLC deliveries on Varian Linacs are achieved through the motion of the leaves from X1 to X2 direction. If the higher dose gradient is at the end of the leaf motion sequence, this can Result in an increased error in the overall dose delivery. Such errors have been observed in Lateral beams for Left-sided and Medial beams for Right-sided treatments. To evaluate and resolve this issue we adopted an approach where the fluence for such beams was geometrically flipped (mirrored) to treat the higher end of the dose gradient first. Results: Using this method, it was possible to deliver the optimized dose map to the area of interest still using only the X1–X2 direction of the leaf motion. The accuracy of this method was tested on different beam as part of our pre-treatment QA program on both Varian delivery systems. With this approach we found that there was significant improvement in delivery accuracy on both 21EX and TrueBeam systems. Beams of initial Gamma index (3% and 3mm) 89–93% were increased to 98–99%. We also observed superior delivery accuracy with TB compared to the 21EX Conclusion: This work demonstrate the need for a delivery sequence option from X2–X1 in situations where the MLC sequence indicates higher dose gradient component is being delivered at the end of the sequence. Results from this work can be considered in the IMRT beam optimization in the treatment planning systems. Further work will be required to establish the application of this approach in clinical setting.

OSTI ID:
22545228
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Improvements in dose accuracy delivered with static-MLC IMRT on an integrated linear accelerator control system
Journal Article · Tue May 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22098840

TH-E-BRE-05: Analysis of Dosimetric Characteristics in Two Leaf Motion Calculator Algorithms for Sliding Window IMRT
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22412395

Monte Carlo based IMRT dose verification using MLC log files and R/V outputs
Journal Article · Sat Jul 15 00:00:00 EDT 2006 · Medical Physics · OSTI ID:20853232