SU-E-T-20: A Correlation Study of 2D and 3D Gamma Passing Rates for Prostate IMRT Plans

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924381· OSTI ID:22545155
 [1]; ;  [1];  [2]
  1. Fox chase Cancer Center, Philadelphia, Pennsylvania (United States)
  2. Sun Yat-sen University Cancer Center, Guangzhou, Guangdong (China)

Purpose: To investigate the correlation between the two-dimensional gamma passing rate (2D %GP) and three-dimensional gamma passing rate (3D %GP) in prostate IMRT quality assurance. Methods: Eleven prostate IMRT plans were randomly selected from the clinical database and were used to obtain dose distributions in the phantom and patient. Three types of delivery errors (MLC bank sag errors, central MLC errors and monitor unit errors) were intentionally introduced to modify the clinical plans through an in-house Matlab program. This resulted in 187 modified plans. The 2D %GP and 3D %GP were analyzed using different dose-difference and distance-toagreement (1%-1mm, 2%-2mm and 3%-3mm) and 20% dose threshold. The 2D %GP and 3D %GP were then compared not only for the whole region, but also for the PTVs and critical structures using the statistical Pearson’s correlation coefficient (γ). Results: For different delivery errors, the average comparison of 2D %GP and 3D %GP showed different conclusions. The statistical correlation coefficients between 2D %GP and 3D %GP for the whole dose distribution showed that except for 3%/3mm criterion, 2D %GP and 3D %GP of 1%/1mm criterion and 2%/2mm criterion had strong correlations (Pearson’s γ value >0.8). Compared with the whole region, the correlations of 2D %GP and 3D %GP for PTV were better (the γ value for 1%/1mm, 2%/2mm and 3%/3mm criterion was 0.959, 0.931 and 0.855, respectively). However for the rectum, there was no correlation between 2D %GP and 3D %GP. Conclusion: For prostate IMRT, the correlation between 2D %GP and 3D %GP for the PTV is better than that for normal structures. The lower dose-difference and DTA criterion shows less difference between 2D %GP and 3D %GP. Other factors such as the dosimeter characteristics and TPS algorithm bias may also influence the correlation between 2D %GP and 3D %GP.

OSTI ID:
22545155
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-422: Correlation Between 2D Passing Rates and 3D Dose Differences for Pretreatment VMAT QA
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369573

SU-E-T-114: Analysis of MLC Errors On Gamma Pass Rates for Patient-Specific and Conventional Phantoms
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22545240

SU-F-T-272: Patient Specific Quality Assurance of Prostate VMAT Plans with Portal Dosimetry
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22648886