Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Development of an ultra-compact CsI/HgI{sub 2} gamma-ray scintillation spectrometer

Conference ·
OSTI ID:225390
; ;  [1];  [2]; ;  [3]
  1. Xsirius, Inc., Camarillo, CA (United States)
  2. Los Alamos National Lab., NM (United States)
  3. UCLA School of Medicine, Los Angeles, CA (United States)

A novel new semiconductor photodetector has been developed which utilizes large mercuric iodide photodetectors coupled to highly optimized CsI(T1) scintillators for gamma ray spectroscopy. With this new detector technology the authors have achieved energy resolution superior to that of any other scintillation detector. Furthermore, gamma probes based on the new HgI{sub 2}/CsI(Tl) detector can be highly miniaturized offering improved portability. A {1/2}-inch diameter HgI{sub 2} photodetector coupled with a {1/2}-inch diameter by {1/2}-inch high right-rectangular scintillator produced energy resolution of 4.58% FWHM for {sup 137}Cs (662 keV). This is perhaps the best result ever reported for room temperature scintillation spectroscopy. Evaluation of a prototype device with similar performance has been conducted at Los Alamos using Pu and U standard samples. Recently, Monte-Carlo simulations have been performed for co-optimization of the gamma-collection efficiency and light collection efficiency of the scintillator/photodetector pairs resulting in a new tapered scintillator geometry. Energy resolution of 5.69% FWHM at 662 keV was obtained for a 1-inch diameter photodetector coupled to a two-inch long conical CsI(Tl) scintillator; with dimensions: 1-inch diameter at the top tapered to 2-inch diameter at the bottom. The long term stability of the technology has been verified. Current efforts to optimize the detectors for specific applications in safeguards and in materials control and accountability are discussed.

DOE Contract Number:
FG03-94ER81687
OSTI ID:
225390
Report Number(s):
CONF-950787--
Country of Publication:
United States
Language:
English

Similar Records

High efficiency CsI(Tl)/HgI{sub 2} gamma ray spectrometers
Journal Article · Tue Aug 01 00:00:00 EDT 1995 · IEEE Transactions on Nuclear Science · OSTI ID:129192

High resolution scintillation spectroscopy with HgI/sub 2/ as the photodetector
Technical Report · Wed Dec 31 23:00:00 EST 1986 · OSTI ID:6758821

Mercuric iodide photodetectors for scintillation spectroscopy