skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Multi-element CZT array for nuclear safeguards applications

Conference ·
OSTI ID:22531446
; ; ;  [1];  [2]
  1. Korea Institute of Nuclear Nonproliferation and Control, 1534 Yuseong-daero, Yusung-gu, Daejeon 305-348, (Korea, Republic of)
  2. NuCare Medical Systems, Inc., 30 Songdo Miraero, Yeonsu-gu, Incheon, 406-840, (Korea, Republic of)

To address drawbacks of a CZT detector due to its limited size, a dual-CZT array, consisting of two CZT detectors and a DAS (Data Acquisition System), was constructed and algorithm to handle the spectra from the CZT array was discussed in this study. Spectra of 4.5% enriched UO{sub 2} powder were obtained using the dual-CZT array. The detection efficiency acquired by the CZT array was about twice higher than that of a single CZT spectra, while FWHMs are nearly identical (about 7.0 keV) for each spectrum. The results of this study show that a much better detection efficiency of CZT can be achieved without degrading its energy resolution. (authors)

Research Organization:
Institute of Electrical and Electronics Engineers - IEEE, 3 Park Avenue, 17th Floor, New York, N.Y. 10016-5997 (United States)
OSTI ID:
22531446
Report Number(s):
ANIMMA-2015-IO-57; TRN: US16V0218102387
Resource Relation:
Conference: ANIMMA 2015: 4. International Conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications, Lisboa (Portugal), 20-24 Apr 2015; Other Information: Country of input: France; 5 Refs.
Country of Publication:
United States
Language:
English