A new cubic phantom for PET/CT dosimetry: Experimental and Monte Carlo characterization
- Departamento de Ensino, Instituto Federal de Educacao, Ciencia e Tecnologia da Bahia, Campus Vitoria da Conquista, Zabele, Av. Amazonas 3150, 45030-220 Vitoria da Conquista, BA (Brazil)
- Departamento de Fisica, Universidade Federal de Sergipe-UFS, Sao Cristovao, Sergipe (Brazil)
- Instituto de Pesquisas Energeticas e Nucleares - IPEN-CNEN/SP, Av. Prof. Lineu Prestes, 2242, Cidade Universitaria, 05508-000 Sao Paulo SP (Brazil)
In recent years, positron emission tomography (PET) associated with multidetector computed tomography (MDCT) has become a diagnostic technique widely disseminated to evaluate various malignant tumors and other diseases. However, during PET/CT examinations, the doses of ionizing radiation experienced by the internal organs of patients may be substantial. To study the doses involved in PET/CT procedures, a new cubic phantom of overlapping acrylic plates was developed and characterized. This phantom has a deposit for the placement of the fluorine-18 fluoro-2-deoxy-D-glucose ({sup 18}F-FDG) solution. There are also small holes near the faces for the insertion of optically stimulated luminescence dosimeters (OSLD). The holes for OSLD are positioned at different distances from the {sup 18}F-FDG deposit. The experimental results were obtained in two PET/CT devices operating with different parameters. Differences in the absorbed doses were observed in OSLD measurements due to the non-orthogonal positioning of the detectors inside the phantom. This phantom was also evaluated using Monte Carlo simulations, with the MCNPX code. The phantom and the geometrical characteristics of the equipment were carefully modeled in the MCNPX code, in order to develop a new methodology form comparison of experimental and simulated results, as well as to allow the characterization of PET/CT equipments in Monte Carlo simulations. All results showed good agreement, proving that this new phantom may be applied for these experiments. (authors)
- Research Organization:
- Institute of Electrical and Electronics Engineers - IEEE, 3 Park Avenue, 17th Floor, New York, N.Y. 10016-5997 (United States)
- OSTI ID:
- 22531381
- Report Number(s):
- ANIMMA-2015-IO-333; TRN: US16V0361102322
- Resource Relation:
- Conference: ANIMMA 2015: 4. International Conference on Advancements in Nuclear Instrumentation Measurement Methods and their Applications, Lisboa (Portugal), 20-24 Apr 2015; Other Information: Country of input: France; 25 Refs.
- Country of Publication:
- United States
- Language:
- English
Similar Records
Monte Carlo simulations of adult and pediatric computed tomography exams: Validation studies of organ doses with physical phantoms
Correction for FDG PET dose extravasations: Monte Carlo validation and quantitative evaluation of patient studies
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
ABSORBED RADIATION DOSES
COMPARATIVE EVALUATIONS
COMPUTERIZED SIMULATION
DOSEMETERS
DOSIMETRY
FLUORINE 18
GLUCOSE
LUMINESCENCE
M CODES
MONTE CARLO METHOD
NEOPLASMS
ORGANS
PATIENTS
PHANTOMS
PLATES
POSITIONING
POSITRON COMPUTED TOMOGRAPHY