SU-E-T-571: Newly Emerging Integrated Transmission Detector Systems Provide Online Quality Assurance of External Beam Radiation Therapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924933· OSTI ID:22496285
; ; ; ;  [1]
  1. UC Davis Cancer Center, Sacramento, CA (United States)
Purpose: Two newly emerging transmission detectors positioned upstream from the patient have been evaluated for online quality assurance of external beam radiotherapy. The prototype for the Integral Quality Monitor (IQM), developed by iRT Systems GmbH (Koblenz, Germany) is a large-area ion chamber mounted on the linac accessory tray to monitor photon fluence, energy, beam shape, and gantry position during treatment. The ion chamber utilizes a thickness gradient which records variable response dependent on beam position. The prototype of Delta4 Discover™, developed by ScandiDos (Uppsala, Sweden) is a linac accessory tray mounted 4040 diode array that measures photon fluence during patient treatment. Both systems are employable for patient specific QA prior to treatment delivery. Methods: Our institution evaluated the reproducibility of measurements using various beam types, including VMAT treatment plans with both the IQM ion chamber and the Delta4 Discover diode array. Additionally, the IQM’s effect on photon fluence, dose response, simulated beam error detection, and the accuracy of the integrated barometer, thermometer, and inclinometer were characterized. The evaluated photon beam errors are based on the annual tolerances specified in AAPM TG-142. Results: Repeated VMAT treatments were measured with 0.16% reproducibility by the IQM and 0.55% reproducibility by the Delta4 Discover. The IQM attenuated 6, 10, and 15 MV photon beams by 5.43±0.02%, 4.60±0.02%, and 4.21±0.03% respectively. Photon beam profiles were affected <1.5% in the non-penumbra regions. The IQM’s ion chamber’s dose response was linear and the thermometer, barometer, and inclinometer agreed with other calibrated devices. The device detected variations in monitor units delivered (1%), field position (3mm), single MLC leaf positions (13mm), and photon energy. Conclusion: We have characterized two new transmissions detector systems designed to provide in-vivo like measurements upstream from the patient. Both systems demonstrate substantial utility for online treatment verification and QA of photon external beam radiotherapy.
OSTI ID:
22496285
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-F-T-326: Diode Array Transmission Detector Systems Evaluation
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22648932

SU-E-T-354: Efficient and Enhanced QA Testing of Linear Accelerators Using a Real-Time Beam Monitor
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22548404

A method for online verification of adapted fields using an independent dose monitor
Journal Article · Mon Jul 15 00:00:00 EDT 2013 · Medical Physics · OSTI ID:22121632