Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

SU-E-T-538: Lung SBRT Dosimetric Comparison of 3D Conformal and RapidArc Planning

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924900· OSTI ID:22496254
; ;  [1]
  1. Grand River Hospital, Kitchener, ON (Canada)
Purpose: Dose distributions of RapidArc Plan can be quite different from standard 3D conformal radiation therapy. SBRT plans can be optimized with high conformity or mimic the 3D conformal treatment planning with very high dose in the center of the tumor. This study quantifies the dosimetric differences among 3D conformal plan; flattened beam and FFF beam RapidArc Plans for lung SBRT. Methods: Five lung cancer patients treated with 3D non-coplanar SBRT were randomly selected. All the patients were CT scanned with 4DCT to determine the internal target volume. Abdominal compression was applied to minimize respiratory motion for SBRT patients. The prescription dose was 48 Gy in 4 fractions. The PTV coverage was optimized by two groups of objective function: one with high conformity, another mimicking 3D conformal dose distribution with high dose in the center of PTV. Optimization constraints were set to meet the criteria of the RTOG-0915 protocol. All VMAT plans were optimized with the RapidArc technique using four full arcs in Eclipse treatment planning system. The RapidArc SBRT plans with flattened 6MV beam and 6MV FFF beam were generated and dosimetric results were compared with the previous treated 3D non-coplanar plans. Results: All the RapidArc plans with flattened beam and FFF beam had similar results for the PTV and OARs. For the high conformity optimization group, The DVH of PTV exhibited a steep dose fall-off outside the PTV compared to the 3D non-coplanar plan. However, for the group mimicking the 3D conformal target dose distribution, although the PTV is very similar to the 3D conformal plan, the ITV coverage is better than 3D conformal plan. Conclusion: Due to excellent clinical experiences of 3D conformal SBRT treatment, the Rapid Arc optimization mimicking 3D conformal planning may be suggested for clinical use.
OSTI ID:
22496254
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Sci—Sat AM: Stereo — 06: Dosimetric Comparison of 3D Conformai, Flattened and Flattening Filter-Free TrueBeam RapidArc Planning for Lung SBRT
Journal Article · Fri Aug 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22409500

Poster — Thur Eve — 31: Dosimetric Effect of Respiratory Motion on RapidArc Lung SBRT Treatment Delivered by TrueBeam Linear Accelerator
Journal Article · Fri Aug 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22407654

Fast Arc Delivery for Stereotactic Body Radiotherapy of Vertebral and Lung Tumors
Journal Article · Tue May 01 00:00:00 EDT 2012 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22056359