SU-E-J-140: Simulation Study Using Thermoacoustics to Image Proton Dose and Range in Water and Skull Phantom
- Purdue University, West Lafayette, IN (United States)
- St. Jude Children’s Research Hospital, Memphis, TN (United States)
Purpose: In this study, thermoacoustic pressure signals generated from a proton beam were simulated in water and currently within a skull phantom to investigate the sensitivity of radioacoustic CT imaging in the brain. Methods: Thermoacoustically generated pressure signals from a pulse pencil proton beam (12, 15, 20, and 27cm range) were simulated in water. These simulated pressure signal are detected using a (71) transducer array placed along the surface of a cylinder (30cm × 40cm) and rotated over 2π (in 2 degree increments), where the normal vector to the surface of each transducer intersects the isocenter of the scanner. Currently, a software skull phantom is positioned at isocenter, where the scattering, absorption and speed of dispersion of the thermoacoustic signal through a three layer cortical-trabecular-cortical structure is being simulated. Based on data obtained from the literature, the effects of acoustic attenuation and speed-of-sound (dispersion) will be applied within the 3D FBP algorithm to obtain dosimetric images. Results: Based on hydrophone detector specifications, a 0.5MHz bandwidth and 50dB re 1μPa per Hz^1/2, a 1.6cGy sensitivity at the Bragg peak was demonstrated while maintaining a 1.0 mm (FWHM) range resolution along the central axis of the beam. Utilizing this same information, the integral dose within the Bragg peak and distal edge compared to MC had a 2% (statistical) and 5% voxel-based RMS at this same dose sensitivity. We plan to present preliminary data determining the range sensitivity for a head phantom for this scanner design and the feasibility of imaging the proton dose in patients with a brain tumor undergoing therapy. Conclusion: RACT scanner provides 3D dosimetric images with 1.6cGy (Bragg peak) sensitivity with 1mm range sensitivity. Simulations will be performed to determine feasibility to treat brain cancer patients.
- OSTI ID:
- 22494152
- Journal Information:
- Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Feasibility of RACT for 3D dose measurement and range verification in a water phantom
Thermoacoustic range verification in the presence of acoustic heterogeneity and soundspeed errors - Robustness relative to ultrasound image of underlying anatomy