SU-E-J-42: Evaluation of Fiducial Markers for Ultrasound and X-Ray Images Used for Motion Tracking in Pancreas SBRT

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924129· OSTI ID:22494066
; ; ; ; ;  [1];  [2]; ; ;  [3]
  1. Department of Radiation Oncology, Johns Hopkins University, Baltimore, MD (United States)
  2. Department of Mechanical Engineering, Johns Hopkins University, Baltimore, MD (United States)
  3. Department of Computer Science, Johns Hopkins University, Baltimore, MD (United States)

Purpose Ultrasound tracking of target motion relies on visibility of vascular and/or anatomical landmark. However this is challenging when the target is located far from vascular structures or in organs that lack ultrasound landmark structure, such as in the case of pancreas cancer. The purpose of this study is to evaluate visibility, artifacts and distortions of fusion coils and solid gold markers in ultrasound, CT, CBCT and kV images to identify markers suitable for real-time ultrasound tracking of tumor motion in SBRT pancreas treatment. Methods Two fusion coils (1mm × 5mm and 1mm × 10 mm) and a solid gold marker (0.8mm × 10mm) were embedded in a tissue–like ultrasound phantom. The phantom (5cm × 12cm × 20cm) was prepared using water, gelatin and psyllium-hydrophilic-mucilloid fiber. Psylliumhydrophilic mucilloid acts as scattering medium to produce echo texture that simulates sonographic appearance of human tissue in ultrasound images while maintaining electron density close to that of water in CT images. Ultrasound images were acquired using 3D-ultrasound system with markers embedded at 5, 10 and 15mm depth from phantom surface. CT images were acquired using Philips Big Bore CT while CBCT and kV images were acquired with XVI-system (Elexta). Visual analysis was performed to compare visibility of the markers and visibility score (1 to 3) were assigned. Results All markers embedded at various depths are clearly visible (score of 3) in ultrasound images. Good visibility of all markers is observed in CT, CBCT and kV images. The degree of artifact produced by the markers in CT and CBCT images are indistinguishable. No distortion is observed in images from any modalities. Conclusion All markers are visible in images across all modalities in this homogenous tissue-like phantom. Human subject data is necessary to confirm the marker type suitable for real-time ultrasound tracking of tumor motion in SBRT pancreas treatment.

OSTI ID:
22494066
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-G-JeP4-12: Real-Time Organ Motion Monitoring Using Ultrasound and KV Fluoroscopy During Lung SBRT Delivery
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649461

Automatic tracking of arbitrarily shaped implanted markers in kilovoltage projection images: A feasibility study
Journal Article · Tue Jul 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22412482

SU-E-J-133: Evaluation of Inter- and Intra-Fractional Pancreas Tumor Residual Motions with Abdominal Compression
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22334028