Title: Numerical study of a confocal ultrasonic setup for creation of cavitation

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.4934399· OSTI ID:22492628

Acoustic cavitation is used for various therapeutic applications such as local enhancement of drug delivery, histotripsy or hyperthermia. One of the utmost important parameter for cavitation creation is the rarefaction pressure. The typical magnitude of the rarefaction pressure required to initiate cavitation from gas dissolved in tissue is beyond the range of the megapascal. Because nonlinear effects need to be taken into account, a numerical simulator based on the Westervelt equation was used to study the pressure waveform and the acoustic field generated by a setup for creation of cavitation consisting of two high intensity focused ultrasound transducers mounted confocally. At constant acoustic power, simulations with only one and both transducers from the confocal setup showed that the distortion of the pressure waveform due to the combined effects of nonlinearity and diffraction is less pronounced when both confocal transducers are used. Consequently, the confocal setup generates a greater peak negative pressure at focus which is more favorable for cavitation initiation. Comparison between the confocal setup and a single transducer with the same total emitting surface puts in evidence the role of the spatial separation of the two beams. Furthermore, it has been previously shown that the location of the peak negative pressure created by a single transducer shifts from focus towards the transducers in the presence of nonlinear effects. The simulator was used to study a configuration where the acoustical axes of transducers intersect on the peak negative pressure instead of the geometrical focus. For a representative confocal setup, namely moderate nonlinear effects, a 2% increase of the peak negative pressure and 8% decrease of the peak positive pressure resulted from this configuration. These differences tend to increase by increasing nonlinear effects. Although the optimal position of the transducers varies with the nonlinear regimen, the intersection point remains the location of the peak negative pressure in any case. Thus, unlike the location of the peak negative pressure for a single transducer can shift by a few millimeters, the focal point of a confocal device is independent of the power. This point is particularly important for therapeutic applications, frequently requiring high spatial accuracy. An experiment conducted shows that cavitation creation can be achieved easier with confocal ultrasound.

OSTI ID:
22492628
Journal Information:
AIP Conference Proceedings, Vol. 1685, Issue 1; Conference: 20. international symposium on nonlinear acoustics, Ecully (France), 29 Jun - 3 Jul 2015, 2. international sonic boom forum, Ecully (France), 29 Jun - 3 Jul 2015; Other Information: (c) 2015 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English