Title: Note: An X-ray powder diffractometer with a wide scattering-angle range of 72° using asymmetrically positioned one-dimensional detectors

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4940986· OSTI ID:22482845
;  [1];  [1]; ;  [2];  [1]
  1. Synchrotron X-ray Station at SPring-8, National Institute for Materials Science (NIMS), 1-1-1 Kouto, Sayo, Hyogo 679-5148 (Japan)
  2. Global Research Center for Environment and Energy based Nanomaterials Science (GREEN), Lithium Air Battery Specially Promoted Research Team, NIMS, 1-1 Namiki, Tsukuba, Ibaraki 305-0044 (Japan)

An X-ray powder diffractometer has been developed for a time-resolved measurement without the requirement of a scattering angle (2θ) scan. Six one-dimensional detector modules are asymmetrically arranged in a vertical line at a designed distance of 286.5 mm. A detector module actually covers a diffraction angle of about 12° with an angular resolution of 0.01°. A diffracted intensity pattern is simultaneously recorded in a 2θ angular range from 1.63° to 74.37° in a “one shot” measurement. We tested the performance of the diffractometer with reference CeO{sub 2} powders and demonstrated diffraction measurements from an operating lithium-air battery.

OSTI ID:
22482845
Journal Information:
Review of Scientific Instruments, Vol. 87, Issue 1; Other Information: (c) 2016 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English