Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

The Effect of Realtime Monitoring on Dose Exposure to Staff Within an Interventional Radiology Setting

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2];  [3]; ;  [1]
  1. Baptist Hospital of Miami, Miami Cardiac & Vascular Institute (MCVI) (United States)
  2. Philips HealthCare, Clinical Science Interventional X-ray (Netherlands)
  3. Kantonsspital Aarau, Clinical and Interventional Angiology (Switzerland)
PurposeThe purpose of this study is to evaluate a new device providing real-time monitoring on radiation exposure during fluoroscopy procedures intending to reduce radiation in an interventional radiology setting.Materials and MethodsIn one interventional suite, a new system providing a real-time radiation dose display and five individual wireless dosimeters were installed. The five dosimeters were worn by the attending, fellow, nurse, technician, and anesthesiologist for every procedure taking place in that suite. During the first 6-week interval the dose display was off (closed phase) and activated thereafter, for a 6-week learning phase (learning phase) and a 10-week open phase (open phase). During these phases, the staff dose and the individual dose for each procedure were recorded from the wireless dosimeter and correlated with the fluoroscopy time. Further subanalysis for dose exposure included diagnostic versus interventional as well as short (<10 min) versus long (>10 min) procedures.ResultsA total of 252 procedures were performed (n = 88 closed phase, n = 50 learning phase, n = 114 open phase). The overall mean staff dose per fluoroscopic minute was 42.79 versus 19.81 µSv/min (p < 0.05) comparing the closed and open phase. Thereby, anesthesiologists were the only individuals attaining a significant dose reduction during open phase 16.9 versus 8.86 µSv/min (p < 0.05). Furthermore, a significant reduction of total staff dose was observed for short 51 % and interventional procedures 45 % (p < 0.05, for both).ConclusionA real-time qualitative display of radiation exposure may reduce team radiation dose. The process may take a few weeks during the learning phase but appears sustained, thereafter.
OSTI ID:
22469761
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 5 Vol. 38; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Personalized Feedback on Staff Dose in Fluoroscopy-Guided Interventions: A New Era in Radiation Dose Monitoring
Journal Article · Tue Nov 14 23:00:00 EST 2017 · Cardiovascular and Interventional Radiology · OSTI ID:22756337

SU-G-IeP3-13: Real-Time Patient and Staff Dose Monitoring in Fluoroscopy Guided Interventions
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649406

Radiation dose to personnel during percutaneous renal calculus removal
Journal Article · Sat Nov 30 23:00:00 EST 1985 · AJR, Am. J. Roentgenol.; (United States) · OSTI ID:6034582