Patient-Specific Quality Assurance for the Delivery of {sup 60}Co Intensity Modulated Radiation Therapy Subject to a 0.35-T Lateral Magnetic Field
Journal Article
·
· International Journal of Radiation Oncology, Biology and Physics
Purpose: This work describes a patient-specific dosimetry quality assurance (QA) program for intensity modulated radiation therapy (IMRT) using ViewRay, the first commercial magnetic resonance imaging-guided RT device. Methods and Materials: The program consisted of: (1) a 1-dimensional multipoint ionization chamber measurement using a customized 15-cm{sup 3} cube-shaped phantom; (2) 2-dimensional (2D) radiographic film measurement using a 30- × 30- × 20-cm{sup 3} phantom with multiple inserted ionization chambers; (3) quasi-3D diode array (ArcCHECK) measurement with a centrally inserted ionization chamber; (4) 2D fluence verification using machine delivery log files; and (5) 3D Monte Carlo (MC) dose reconstruction with machine delivery files and phantom CT. Results: Ionization chamber measurements agreed well with treatment planning system (TPS)-computed doses in all phantom geometries where the mean ± SD difference was 0.0% ± 1.3% (n=102; range, −3.0%-2.9%). Film measurements also showed excellent agreement with the TPS-computed 2D dose distributions where the mean passing rate using 3% relative/3 mm gamma criteria was 94.6% ± 3.4% (n=30; range, 87.4%-100%). For ArcCHECK measurements, the mean ± SD passing rate using 3% relative/3 mm gamma criteria was 98.9% ± 1.1% (n=34; range, 95.8%-100%). 2D fluence maps with a resolution of 1 × 1 mm{sup 2} showed 100% passing rates for all plan deliveries (n=34). The MC reconstructed doses to the phantom agreed well with planned 3D doses where the mean passing rate using 3% absolute/3 mm gamma criteria was 99.0% ± 1.0% (n=18; range, 97.0%-100%), demonstrating the feasibility of evaluating the QA results in the patient geometry. Conclusions: We developed a dosimetry program for ViewRay's patient-specific IMRT QA. The methodology will be useful for other ViewRay users. The QA results presented here can assist the RT community to establish appropriate tolerance and action limits for ViewRay's IMRT QA.
- OSTI ID:
- 22423840
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 1 Vol. 91; ISSN IOBPD3; ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
IMRT Quality Assurance Using a Second Treatment Planning System
TH-C-12A-01: Develop a Patient-Specific QA Program for Radiation Therapy with On-Board MRI
WE-D-BRA-07: Analysis of ArcCHECK Diode Array Performance for ViewRay Quality Assurance
Journal Article
·
Thu Dec 31 23:00:00 EST 2009
· Medical Dosimetry
·
OSTI ID:21486870
TH-C-12A-01: Develop a Patient-Specific QA Program for Radiation Therapy with On-Board MRI
Journal Article
·
Sun Jun 15 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22409918
WE-D-BRA-07: Analysis of ArcCHECK Diode Array Performance for ViewRay Quality Assurance
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22570160