TH-E-17A-10: Markerless Lung Tumor Tracking Based On Beams Eye View EPID Images
- UT Southwestern Medical Center, Dallas, Texas (United States)
- University of Texas at Dallas, Richardson, Texas (United States)
- Premier Cancer Centers, Dallas, TX (United States)
Purpose: Dynamic tumor tracking or motion compensation techniques have proposed to modify beam delivery following lung tumor motion on the flight. Conventional treatment plan QA could be performed in advance since every delivery may be different. Markerless lung tumor tracking using beams eye view EPID images provides a best treatment evaluation mechanism. The purpose of this study is to improve the accuracy of the online markerless lung tumor motion tracking method. Methods: The lung tumor could be located on every frame of MV images during radiation therapy treatment by comparing with corresponding digitally reconstructed radiograph (DRR). A kV-MV CT corresponding curve is applied on planning kV CT to generate MV CT images for patients in order to enhance the similarity between DRRs and MV treatment images. This kV-MV CT corresponding curve was obtained by scanning a same CT electron density phantom by a kV CT scanner and MV scanner (Tomotherapy) or MV CBCT. Two sets of MV DRRs were then generated for tumor and anatomy without tumor as the references to tracking the tumor on beams eye view EPID images. Results: Phantom studies were performed on a Varian TrueBeam linac. MV treatment images were acquired continuously during each treatment beam delivery at 12 gantry angles by iTools. Markerless tumor tracking was applied with DRRs generated from simulated MVCT. Tumors were tracked on every frame of images and compared with expected positions based on programed phantom motion. It was found that the average tracking error were 2.3 mm. Conclusion: This algorithm is capable of detecting lung tumors at complicated environment without implanting markers. It should be noted that the CT data has a slice thickness of 3 mm. This shows the statistical accuracy is better than the spatial accuracy. This project has been supported by a Varian Research Grant.
- OSTI ID:
- 22412522
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
TH-AB-202-01: Daily Lung Tumor Motion Characterization On EPIDs Using a Markerless Tiling Model
Feasibility Study for Markerless Tracking of Lung Tumors in Stereotactic Body Radiotherapy
Registration of clinical volumes to beams-eye-view images for real-time tracking
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22679217
Feasibility Study for Markerless Tracking of Lung Tumors in Stereotactic Body Radiotherapy
Journal Article
·
Fri Oct 01 00:00:00 EDT 2010
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:21436206
Registration of clinical volumes to beams-eye-view images for real-time tracking
Journal Article
·
Sun Dec 14 23:00:00 EST 2014
· Medical Physics
·
OSTI ID:22403163