Simulation study of the effects of near- and far-field heating during focused ultrasound uterine fibroid ablation using an electronically focused phased array: A theoretical analysis of patient safety

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4883777· OSTI ID:22412474
;  [1]
  1. Sunnybrook Research Institute, Toronto, Ontario M4N 3M5, Canada and Department of Medical Biophysics, University of Toronto, Toronto, Ontario M5G 1L7 (Canada)

Purpose: Assess the feasibility of using large-aperture, flat ultrasonic transducer arrays with 6500 small elements operating at 500 kHz without the use of any mechanical components for the thermal coagulation of uterine fibroids. This study examines the benefits and detriments of using a frequency that is significantly lower than that used in clinical systems (1–1.5 MHz). Methods: Ultrasound simulations were performed using the anatomies of five fibroid patients derived from 3D MRI. Using electronic steering solely, the ultrasound focus from a flat, 6500-element phased array was translated around the volume of the fibroids in various patterns to assess the feasibility of completing full treatments from fixed physical locations. Successive temperature maps were generated by numerically solving the bioheat equation. Using a thermal dose model, the bioeffects of these simulations were quantified and analyzed. Results: The simulations indicate that such an array could be used to perform fibroid treatments to 18 EM{sub 43} at an average rate of 90 ± 20 cm{sup 3}/h without physically moving the transducer array. On average, the maximum near-field thermal dose for each patient was below 4 EM{sub 43}. Fibroid tissue could be treated as close as 40 mm to the spine without reaching temperatures expected to cause pain or damage. Conclusions: Fibroids were successfully targeted and treated from a single transducer position to acceptable extents and without causing damage in the near- or far-field. Compared to clinical systems, treatment rates were good. The proposed treatment paradigm is a promising alternative to existing systems and warrants further investigation.

OSTI ID:
22412474
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 7 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

MR-Guided Focused Ultrasound for the Treatment of Uterine Fibroids
Journal Article · Thu Feb 14 23:00:00 EST 2013 · Cardiovascular and Interventional Radiology · OSTI ID:22156519

Targeted Vessel Ablation for More Efficient Magnetic Resonance-Guided High-Intensity Focused Ultrasound Ablation of Uterine Fibroids
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Cardiovascular and Interventional Radiology · OSTI ID:22066538

Laparoscopic Ultrasound-Guided Radiofrequency Ablation of Uterine Fibroids
Journal Article · Tue Aug 15 00:00:00 EDT 2006 · Cardiovascular and Interventional Radiology · OSTI ID:21091172