Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

WE-G-18A-02: Calibration-Free Combined KV/MV Short Scan CBCT

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889513· OSTI ID:22409768
; ; ;  [1];  [2]
  1. Stanford University, Stanford, CA (United States)
  2. Varian Medical Systems, Palo Alto, CA (United States)
Purpose: To combine orthogonal kilo-voltage (kV) and Mega-voltage (MV) projection data for short scan cone-beam CT to reduce imaging time on current radiation treatment systems, using a calibration-free gain correction method. Methods: Combining two orthogonal projection data sets for kV and MV imaging hardware can reduce the scan angle to as small as 110° (90°+fan) such that the total scan time is ∼18 seconds, or within a breath hold. To obtain an accurate reconstruction, the MV projection data is first linearly corrected using linear regression using the redundant data from the start and end of the sinogram, and then the combined data is reconstructed using the FDK method. To correct for the different changes of attenuation coefficients in kV/MV between soft tissue and bone, the forward projection of the segmented bone and soft tissue from the first reconstruction in the redundant region are added to the linear regression model. The MV data is corrected again using the additional information from the segmented image, and combined with kV for a second FDK reconstruction. We simulated polychromatic 120 kVp (conventional a-Si EPID with CsI) and 2.5 MVp (prototype high-DQE MV detector) projection data with Poisson noise using the XCAT phantom. The gain correction and combined kV/MV short scan reconstructions were tested with head and thorax cases, and simple contrast-to-noise ratio measurements were made in a low-contrast pattern in the head. Results: The FDK reconstruction using the proposed gain correction method can effectively reduce artifacts caused by the differences of attenuation coefficients in the kV/MV data. The CNRs of the short scans for kV, MV, and kV/MV are 5.0, 2.6 and 3.4 respectively. The proposed gain correction method also works with truncated projections. Conclusion: A novel gain correction and reconstruction method was developed to generate short scan CBCT from orthogonal kV/MV projections. This work is supported by NIH Grant 5R01CA138426-05.
OSTI ID:
22409768
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

A technique for on-board CT reconstruction using both kilovoltage and megavoltage beam projections for 3D treatment verification
Journal Article · Thu Sep 15 00:00:00 EDT 2005 · Medical Physics · OSTI ID:20726232

Metal artifact correction for x-ray computed tomography using kV and selective MV imaging
Journal Article · Sun Dec 14 23:00:00 EST 2014 · Medical Physics · OSTI ID:22403181

SU-E-I-56: Scan Angle Reduction for a Limited-Angle Intrafraction Verification (LIVE) System
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22325160