Title: Independent dosimetric assessment of the model EP917 episcleral brachytherapy plaque

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4892603· OSTI ID:22409598
;  [1]
  1. Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky 40536 (United States)

Purpose: To investigate the influence of slot design on dose distributions and dose-volume histograms (DVHs) for the model EP917 plaque for episcleral brachytherapy. Methods: Dimensions and orientations of the slots were measured for three model EP917 plaques and compared to data in the Plaque Simulator (PS) treatment planning software (version 5.7.6). These independently determined coordinates were incorporated into the MCNP Monte Carlo simulation environment to obtain dose from the plaques in a water environment and in a clinical environment with ocular structures. A tumor volume was simulated as 5 mm in apical height and 11 mm in basal diameter. Variations in plaque mass density and composition; slot length, width, and depth; seed positioning; and Ag-marker rod positioning were simulated to examine their influence on plaque central axis (CAX) and planar dose distributions, and DVHs. Results: Seed shifts in a single slot toward the eye and shifts of the{sup 125}I-coated Ag rod within the capsule had the greatest impact on CAX dose distribution. A shift of 0.0994 mm toward the eye increased dose by 14%, 9%, 4.3%, and 2.7% at 1, 2, 5, and 10 mm, respectively, from the inner sclera. When examining the fully-modeled plaque in the ocular geometry, the largest dose variations were caused by shifting the Ag rods toward the sclera and shifting the seeds away from the globe when the slots were made 0.51 mm deeper, causing +34.3% and −69.4% dose changes to the outer sclera, respectively. At points along the CAX, dose from the full plaque geometry using the measured slot design was 2.4% ± 1.1% higher than the manufacturer-provided slot design and 2.2% ± 2.3% higher than the homogeneous calculation of PS treatment planning results. The ratio of D{sub 10} values for the measured slot design to the D{sub 10} values for the manufacturer-provided slot design was higher by 9%, 10%, and 19% for the tumor, inner sclera, and outer sclera, respectively. In comparison to the measured slot design, a theoretical plaque having narrower and deeper slots delivered 30%, 37%, and 62% lower D{sub 10} doses to the tumor, inner sclera, and outer sclera, respectively. Conclusions: While the measured positions of the slots on the model EP917 plaque were in close agreement (<0.7 mm) with the PS values, small differences in the slot shape caused substantial differences in dose distributions and DVH metrics. Increasing slot depth by 0.1 mm decreased outer scleral dose by 20%, yet shifting the Ag rods in the seeds toward the globe by 0.1 mm increased outer scleral dose by 35%. The clinical medical physicist is advised to measure these types of plaques upon acceptance testing before clinical use to inspect slot shape and position for comparison with data used for treatment planning purposes.

OSTI ID:
22409598
Journal Information:
Medical Physics, Vol. 41, Issue 9; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

WE-A-17A-12: The Influence of Eye Plaque Design On Dose Distributions and Dose- Volume Histograms
Journal Article · Sun Jun 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22409598

SU-E-T-15: A Comparison of COMS and EP917 Eye Plaque Applicators Using Different Radionuclides
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22409598

Palladium-103 versus iodine-125 for ophthalmic plaque radiotherapy
Journal Article · Mon Nov 15 00:00:00 EST 1993 · International Journal of Radiation Oncology, Biology and Physics; (United States) · OSTI ID:22409598