MO-C-17A-05: A Three-Dimensional Head-And-Neck Phantom for Validation of Kilovoltage- and Megavoltage-Based Deformable Image Registration
- UC San Francisco, San Francisco, CA (United States)
Purpose: To develop a three-dimensional (3D) deformable head-and-neck (H and N) phantom with realistic tissue contrast for both kilovoltage and megavoltage computed tomography and use it to objectively evaluate deformable image registration (DIR) algorithms. Methods: The phantom represents H and N patient anatomy. It is constructed from thermoplastic, which becomes pliable in boiling water, and hardened epoxy resin. Using a system of additives, the Hounsfield unit (HU) values of these materials were tuned to mimic anatomy for both kilovoltage (kV) and megavoltage (MV) imaging. The phantom opened along a sagittal midsection to reveal nonradiopaque markers, which were used to characterize the phantom deformation. The deformed and undeformed phantom was scanned with kV and MV computed tomography. Additionally, a calibration curve was created to change the HUs of the MV scans to be similar to kV HUs, (MC). The extracted ground-truth deformation was then compared to the results of two commercially available DIR algorithms, from Velocity Medical Solutions and MIM Software. Results: The phantom produced a 3D deformation, representing neck flexion, with a magnitude of up to 8 mm and was able represent tissue HUs for both kV and MV imaging modalities. The two tested deformation algorithms yielded vastly different results. For kV-kV registration, MIM made the lowest mean error, and Velocity made the lowest maximum error. For MV-MV, kV-MV, and kV-MC Velocity produced both the lowest mean and lowest maximum errors. Conclusion: The application of DIR across different imaging modalities is particularly difficult, due to differences in tissue HUs and the presence of imaging artifacts. For this reason, DIR algorithms must be validated specifically for this purpose. The developed H and N phantom is an effective tool for this purpose.
- OSTI ID:
- 22407761
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
A three-dimensional head-and-neck phantom for validation of multimodality deformable image registration for adaptive radiotherapy
The need for application-based adaptation of deformable image registration
SU-E-J-89: Comparative Analysis of MIM and Velocity’s Image Deformation Algorithm Using Simulated KV-CBCT Images for Quality Assurance
Journal Article
·
Sun Dec 14 23:00:00 EST 2014
· Medical Physics
·
OSTI ID:22403169
The need for application-based adaptation of deformable image registration
Journal Article
·
Mon Jan 14 23:00:00 EST 2013
· Medical Physics
·
OSTI ID:22099164
SU-E-J-89: Comparative Analysis of MIM and Velocity’s Image Deformation Algorithm Using Simulated KV-CBCT Images for Quality Assurance
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22494107