Note: O-ring stack system for electron gun alignment
Journal Article
·
· Review of Scientific Instruments
- Korea Research Institute of Standards and Science (KRISS), Daejeon 305-340 (Korea, Republic of)
We present a reliable method for aligning an electron gun which consists of an electron source and lenses by controlling a stack of rubber O-rings in a vacuum condition. The beam direction angle is precisely tilted along two axes by adjusting the height difference of a stack of O-rings. In addition, the source position is shifted in each of three orthogonal directions. We show that the tilting angle and linear shift along the x and y axes as obtained from ten stacked O-rings are ±2.55° and ±2 mm, respectively. This study can easily be adapted to charged particle gun alignment and adjustments of the flange position in a vacuum, ensuring that its results can be useful with regard to electrical insulation between flanges with slight modifications.
- OSTI ID:
- 22392340
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 1 Vol. 86; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English